Cancellation of the actual zero flow rate setting:
Menu: Zero setting, press the
The display will read "Setting Completed" and show the flow rate as indicated by the meter prior to entry to
the configuration mode. Press the
will remain unchanged. Press the
"Zero Setting" item selected. The cancellation of zero setting will take effect upon leaving the meter
configuration mode. The three flashing exclamation marks on the flow rate display will disappear.

7.1.10. 100 per cent

Default value of the "100 Percent" quantity is Q
and maximum flow rates in section 4.1.1.). This default value can be reset, in particular in cases where the
actual maximum flow rate is lower than Q
Menu: 100 percent, press the
The display will read "100 Percent xxxx". Using the
and increase/decrease the selected figure by pressing the
number equal to the flow rate identified with 100%. Press the
Press Any Key". Press any key, preferably the
display.

7.1.11. Exit

Select "Exit" to leave the configuration mode and protect the meter from any unauthorised configuration
action.
Menu: Exit, press the
EXIT
New password
Browse through the menu using the
item appears on the first line with the first character blinking. Select "Exit" and press the
will read "Write To EEPROM Press Any Key". Press any key, preferably the
mode will be terminated, and the data display menu will appear on the display. Use the
browse downwards and upwards through the menu items.
Select "New password" to cancel the existing password and define a new password enabling entry into the
Parameter Setting mode.
Menu: Exit, press the
The display will read "Password 0000". Using the
increase/decrease the selected figure by pressing the
and press the
key. The display will read "Value Entered Press Any Key". Press any key, preferably the
key. The Exit menu will reappear on the display. Select the Exit item and press the
read "Write To EEPROM Press Any Key". Press any key, preferably the
the configuration mode and bring the data display menu on the meter display. Unless the configuration
procedure is terminated in this way, the flowmeter parameters will not be protected by a password. Browse
through the menu items using the
Warning!
After downloading new data to EEPROM, the flowmeter mustn't be turned off for at least 5 minutes!!!
ELIS PLZEŇ a. s., Luční 425/15, 301 00 Plzeň, Czech Republic, Phone: +420/377 517 711, Fax: +420/377 517 722
Design, Assembly and Service Manual
Electromagnetic flowmeter FLONET FN20xx.1
key
key . The display will read "Setting Cancelled" while the flow rate value
key. The basic configuration menu will reappear on the display with the
.
4
key
key
(direction downwards) and
key / New password, press the
(direction downwards) and
as specified for the sensor DN (see the table of minimum
4
key, move the cursor to the desired decimal position
and
keys, respectively. Replace xxxx by a
key. The display will read "Value Entered
key. The basic configuration menu will reappear on the
key
key, move the cursor to the desired decimal position and
and
keys, respectively. Enter the new password
keys (upwards).
Page 45 of 69
keys (upwards). The selected menu
key. The display
key. The meter configuration
and
ley. The display will
key. This command will terminate
keys to
Es90420K/c

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents