Figure 1-3: Factory setting
P: system pressure
MAW: initial value of the measuring range (Lower range limit [LRL])
MEW: final value of the measuring range (Upper range limit [URL])
1: OU2 = I
2: OU2 = InEG
In the set measuring range the output signal is between 4 and 20 mA (OU2 = I) or between
20 and 4 mA (OU2 = InEG). When pressure is outside the set measuring range, the output
signal is as follows:
System pressure above the measuring range:
System pressure below the measuring range:
1.3.3
Customer-specific calibration (sensor trim)
A customer specific calibration (sensor trim) corrects the pressure offset and pressure
range to match a pressure standard. This provides a slope correction to the transmitter's
current characterization curve. An accurate pressure standard is required for a full
calibration. For more information on performing a Sensor Trim, see
Sensor Trim is a two-point sensor calibration that applies two end-point pressures
(CP1, CP2) then linearizes the output between them. CP1 and CP2 must be within the
scaled measuring range. For more information, see
Zero trim is a single-point offset adjustment. It compensates for mounting position
effects and is most effective when performed with the transmitter installed in its final
Quick Start Guide
Output signal > 20 mA if OU2 = I.
Output signal 4 to 3.8 mA if OU2 = InEG.
Output signal 4 to 3.8 mA if OU2 = I.
Output signal > 20 mA if OU2 = InEG.
Figure 1-4: Transmitter reranged
Perform a sensor
Analog
signal.
Introduction
January 2019
trim.