Baker SterilGARD III Advance Operator's Manual page 30

Class ii, type a2 biosafety cabinet
Table of Contents

Advertisement

Document: 333D012G
3. Slide the new exhaust filter in all the way in until it stops against the rear filter stop. Be careful not to contact
the filter media with your fingers. Align the filter frame, left to right, so it is positioned directly over the exhaust
plenum.
4. Tighten the two (2) ½" hex head exhaust plenum clamping bolts, alternating two or three revolutions at a time
until the filter gasket is compressed approximately 1/8" [3.2mm].
5. Tighten the two (2) ½" hex head down flow plenum clamping bolts, alternating two or three revolutions at a
time until the filter gasket is compressed approximately 1/8" [3.2mm].
6. Reassemble the cabinet following the previous instructions in reverse order.
When reinstalling the front seal panel, first tighten each ¼-20 hex cap screw until the bolt seal washer is engaged
with the panel. Next turn each bolt 1 ¼ to 2 ½ revolutions until the panel gasket is evenly compressed
approximately half way. Do not over tighten. Use caution if using a power driven tool.
7. Reconnect power.
8. Check the filter and gasket for leaks.
IMPORTANT
- 29 -
5-18-2007