Appendix A - Caen A 1732 Technical Information Manual

High voltage boards
Table of Contents

Advertisement

Document type:
Title:
User's Manual (MUT)
Mod. A1732-A1832 High Voltage Boards
Notes for Mod. A1832 E
Each A1832 E channel is provided with an "HV EN" input (on the right side of the SHV
connector) that disables the channel when the two pins are not shorted (jumper
removed).
When the channel is disabled, the voltage output (if ON) drops to zero at the maximum
rate available. If the jumper is removed the channel remains OFF and any attempt to turn
the channel ON will result unsuccessful.
NPO:
00101/00:17XXx.MUTx/05

APPENDIX A

HV EN
ch. i
1 kΩ
100 nF
Fig. A.1 – Mod. A1832 E channel individual enable
Filename:
A1732B-A1832B_REV5.DOC
Revision date:
02/05/2005
upper
pin
VCC
100 kΩ
to CONTROL
LOGIC (TTL)
0=ENABLED