4.22
Connecting the separation column
930 Compact IC Flex Oven SeS/PP/Deg (2.930.2560)
Connecting the guard column
Rinsing the guard column
1 Rinsing the guard column
Place a beaker under the guard column's outlet.
■
Start manual control in MagIC Net and select the high-pressure
■
pump: Manual ▶ Manual control ▶ Pump
– Flow: in accordance with column leaflet
– On
Rinse the guard column with eluent for approx. 5 minutes.
■
Stop the high-pressure pump in the manual control in MagIC Net
■
again: Off.
CAUTION
New separation columns are filled with a solution and sealed with stop-
pers on both sides. Before inserting the column, ensure that this solu-
tion can be mixed with the eluent being used (follow the information
provided by the manufacturer).
NOTE
Connect the separation column only after the initial start-up of the
instrument. Until that point, insert a coupling (6.2744.040) instead of
the guard column and separation column.
4 Installation
■■■■■■■■
35