Problem
Pump battery does not
maintain charge for a
full day.
The fob does not
communicate with the
pump. The LimbLogic
VS logo appears and
is followed by the No
Radio Connection
icon.
Action
• Ensure that the system is properly sealed.
• Ensure that the battery has been charged for
at least 4 hours.
• The pump may have ben exposed to a large
static discharge. Plug in the charger and
press the reset button to reset the pump. The
charger does not need to be plugged into an
outlet.
• Ensure that your socket does not leak
excessively.
• Make sure the pump is on by pressing the
on/off button once and verify that the pump
vibrates in recognition of the button activation.
Wait one minute for the fob screen to turn
off. Press the fob center button again.
• The fob may be too close or too far from the
pump. Place the fob about 2 feet away from
the pump, and press the fob center button
again.
• The pump's internal battery may not be
charged. Charge the pump with the charger
for at least 30 minutes and try again.
• Does the patient have the correct fob-pump
combination? The fob and pump are
electronically matched during the production
process. Contact Ohio Willow Wood with the
serial number of the fob and serial number of
the pump to verify that the set is matched.
• The pump may have ben exposed to a large
static discharge. Plug in the charger and
press the reset button to reset the pump. The
charger does not need to be plugged into an
outlet.
21
Overview
continued on next page