Configuration Screen - OHIO WILLOW WOOD LimbLogic VS Series Prosthetist's Manual

Table of Contents

Advertisement

Configuration Screen

The system is shipped with the User Maximum set at 20 in-Hg and
the Vacuum Control Range set at 6 in-Hg. If these settings are not
sufficient for your patient, you may adjust them as follows:
With the Information Screen displayed, press the left and right
buttons simultaneously.
Use the right and left buttons to adjust the User Maximum.
The User Maximum cannot be set higher than 20 in-hg.
Press the up or down button to scroll to the next screen.
Use the right and left buttons to adjust the Vacuum Control
Range. The Vacuum Control Range cannot be set lower
than 4 in-hg.
Note: Increasing the Vacuum Control Range and decreasing the User
Maximum setting may reduce pump cycling and may increase pump
battery life.
Press the center button to exit the Configuration Screen and
save all current settings.
After the warning screen flashes,
you will see the Configuration Screen.
In this example,
the User Maximum
is set at 18 in-Hg
In this example,
the Vacuum Control Range
is set at 4 in-Hg
15