No Radio Connection; Low Battery Mode; Turning Off The Vacuum Pump - OHIO WILLOW WOOD LimbLogic VS Series Prosthetist's Manual

Table of Contents

Advertisement

No Radio Connection

If the No Radio Connection indicator appears:
• make sure the Vacuum Pump has been turned on
• make sure the Vacuum Pump battery has been charged
• make sure the fob is within 3 feet of the Vacuum Pump
• make sure the fob is more than 3 inches away from the
Vacuum Pump
If the No Radio Connection indicator still appears, contact
Ohio Willow Wood.
Low-Battery Mode
If the Vacuum Pump battery has less than 15% of its capacity
remaining, the Vacuum Pump enters a "stutter" mode.
(Stutter mode is when the pump quickly turns on and off when pulling
vacuum.) Charge as soon as possible.
The fob will display the battery charging indicator when the
unit is plugged into the charger.

Turning off the Vacuum Pump

Press the on/off button on the Vacuum Pump two times. The Vacuum
Pump will vibrate twice.
If the current vacuum level is near the current Upper Set Point, the
Vacuum Pump may skip the vibrations to prevent the vacuum level
from exceeding the Upper Set Point.
17