Levelogger User Guide - Software Version 3.2.3 or Higher
5 Levelogger Settings
After the user starts the Levelogger V3.2.3 or higher Software, the Levelogger
settings window will open.
Click the
5.1 Levelogger Information Settings
The
Levelogger
identification, Project ID and Location fields, sampling settings, altitude and
fluid density input fields. The following is a description of each of the fields:
•
Project ID, input your own Levelogger identification system.
The Project ID is limited to 32 characters.
•
Location, input specific site / location information. The location
is limited to 32 characters.
•
Altitude in feet or meters above sea level, at which the logger is actually
deployed, is input in the altitude field. Water column equivalent pressure
decreases with altitude at a rate of approximately 1.2:1000 in the lower
atmosphere below 5000 m. You can compensate for this by entering
an elevation between -1000 ft below sea level and 16,500 ft
(or -300 m and 5000 m) above sea level. The readings will then be
automatically compensated for elevation. See Section 11.1.3 for more
information on how the Levelogger and Barologger adjust for altitude.
F F i i g g u u r r e e 5 5 - - 1 1 L L e e v v e e l l o o g g g g e e r r S S e e t t t t i i n n g g s s W W i i n n d d o o w w
button to retrieve the current settings from the connected Levelogger.
Information
Settings
window
includes
Levelogger
Page 18