1.0 Introduction 1.1 Solinst Readout Unit (SRU) The Solinst Readout Unit (SRU) is an in-field device designed to take and display instant water level readings from a deployed Solinst datalogger. When connected, you can also select a real-time sampling rate and start a real-time monitoring session independent of the datalogger’s internal logging mode.
SRU User Guide 1.3 Compatibility The SRU is compatible with Solinst dataloggers using the following minimum firmware versions, or higher: Datalogger Firmware Version Levelogger 5 1.004 Barologger 5 1.004 Levelogger 5 Junior 1.004 Levelogger 5 LTC 1.004 Rainlogger 5 1.004 LevelVent 5 1.004...
SRU User Guide 2.0 SRU Operation 2.1 Start Up Press and hold any arrow button for two seconds to turn the SRU on. When starting, the SRU will perform a self-test. The start-up screen is shown first; following a short pause the main menu will be displayed. The SRU has an auto-off after 2 minutes of inactivity.
12 mins Sample Rate: 3/11 Files 3/11 Files Idle Idle Compensation: 1081225-01.10.21-12.30 D 1081225-01.10.21-12.30 D Download Logger Download Logger 1081225-01.15.21-15.56 C 1081225-01.15.21-15.56 C Back Start View Files View Files SRU User Guide 1081225-01.16.21-22.50 U 1081225-01.16.21-22.50 U Home Home Next 3 Next 3 Back Back...
1081225-01.10.21-12.30 D 1081225-01.10.21-12.30 D 1081225-01.10.21-12.30 D Download Logger Download Logger Download Logger 1081225-01.15.21-15.56 C 1081225-01.15.21-15.56 C 1081225-01.15.21-15.56 C View Files View Files View Files 1081225-01.16.21-22.50 U 1081225-01.16.21-22.50 U 1081225-01.16.21-22.50 U SRU User Guide Home Home Next 3 Home Next 3 Next 3 Back Back...
SRU User Guide 3.0 Connecting Dataloggers 3.1 Levelogger Connection The SRU cable connects to the top of a Levelogger’s Direct Read Cable. Hold the Direct Read Cable connection and thread the SRU cable connector onto the Direct Read Cable. Figure 3-1 Levelogger Connection Note: An L5 Threaded or Slip Fit Adaptor can also be used to connect a Levelogger to the SRU cable.
SRU User Guide 4.0 Real-Time Readings 4.1 Take a Reading Select Take a Reading to instantly display the current water level (water column height above the pressure sensor), temperature, and conductivity being detected by the connected datalogger. Note: Only LTC Leveloggers will display conductivity readings. Note: See the Levelogger User Guide (Section 1.2) and Vented Dataloggers User Guide (Section 1.2) for an explanation of the pressure sensor measurement fundamentals.
3) The Software Installation Wizard will guide you through the remaining installation process. Figure 7-1 shows the Levelogger Installation Wizard. 4) Restart the computer after installation is completed. Default Directory is C:\Program Files\Solinst\Levelogger4_6 Figure 7-1 Software Installation Wizard Page 15...
To start the Levelogger Software, click , or click the Start button and select: Programs > Solinst > Levelogger 4 > Levelogger 4.6.1 Go to the SRU tab and select the appropriate Com Port for the connected SRU from the centre drop-down menu and Click the ‘Retrieve SRU Settings’...
Page 17
SRU User Guide Select the file(s) from the list that you want to download, and click the ‘Download Data’ icon Figure 7-4 Levelogger Software SRU Tab Page 17...
SRU User Guide 7.3 View Data Data files can be viewed in the Data Control tab by selecting Finish and Open all, or select Finish if you do not need to view the data right away. Figure 7-5 Finish Data Download Page 18...
Note: To synchronize to a different time/ time zone you first must adjust the computer time using the Windows Operating System Date and Time Settings menu. Then re-start the Solinst Levelogger Software and synchronize your SRU to the new time.
SRU User Guide 8.0 Maintenance 8.1 Battery Installation The SRU uses one 9V replaceable alkaline or lithium battery. To install or replace the battery, remove the SRU from the protective outer case. Push up to remove the battery enclosure door from the back of the SRU. Connect the battery lead to the 9V battery and install the battery in the housing.
Make sure to unzip the firmware file after you have downloaded it, so you can access the *.ssf file. To upload new firmware to a SRU, follow these steps: 1) Open the Solinst Firmware Upgrade Utility from the Utilities menu in the main Levelogger Software. Select the Com Port that the SRU is connected from the drop-down menu.