Power Requirements - Agilent Technologies 7700 Series Site Preparation Manual

Hide thumbs Also See for 7700 Series:
Table of Contents

Advertisement

Power Requirements

You are responsible for providing appropriate electrical power and power
outlets for all of the components in your 7700 ICP- MS system. Power
considerations include voltage ranges of major components, power
configurations, power plugs and cords, water circulator requirements, and
general power requirements.
Voltage Ranges of Major Components
The 7700 ICP- MS mainframe requires a single- phase 30A circuit.
Table 8
lists
the voltage ranges and power requirements for the 7700 ICP- MS and related
equipment. Extra power capacity for future additions is strongly recommended.
Each product listed requires a dedicated circuit. The 7700 ICP-MS and data system should
N O T E
each have a separate circuit breaker.
The 7700 ICP- MS system includes a wide- range power supply that can operate
without reconfiguration, however the foreline pump must be configured for
either of two voltage ranges of single- phase alternating current (ac):
200- 219Vac, 50/60Hz (typical for the U.S. and Japan)
220- 240Vac, 50/60Hz (typical for Europe)
The foreline pump will be configured according to the standard voltage in the
country where the instrument is ordered.
If an instrument is being ordered from one location, but is to be installed in another
C A U T I O N
location with different electrical power characteristics, a special note must be made
on the order that the electrical power at the site is different from the standard
electrical power in that country.
G3280-90010 (Rev.B) Agilent 7700 Series ICP-MS Site Preparation Guide