Ancar SD300 User Manual

Dental chair with silent, automatic and highly reliable movement
Hide thumbs Also See for SD300:

Advertisement

Touch Expert Scandinavian
SD 300
USER MANUAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SD300 and is the answer not in the manual?

Questions and answers

Summary of Contents for Ancar SD300

  • Page 1 Touch Expert Scandinavian SD 300 USER MANUAL...
  • Page 3 Antoni Carles, S.A. Volta dels Garrofers, 41-42 T. (34) 93 754 07 97 Pol. Ind. Els Garrofers F. (34) 93 759 26 04 08340-Vilassar de Mar calidad@ancar-online.com ISO 9001 ISO 13485 (Barcelona-SPAIN) www.ancar-dental.com 109037-AQ-IBE 112630-AQ-IBE EC Declaration of Conformity (Directive 93/42/EEC) Manufacturer’s Name...
  • Page 5: Table Of Contents

    USER MANUAL DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian INDEX DESCRIPTION PAGE 1.- WARNINGS 2.- GUARANTEE 3.- IDENTIFICATION 4.- PRECAUTIONS 5.- CLAUSES 6.- TECHNICAL FEATURES 6.1.- Chair 6.2.- Dental unit 7.- TRANSPORT AND DIMENSIONS 8.- COMPOSITION 8.1.- Circuit box 8.2.- Instrument control pedal (proportional) 8.3.- Instrument tray 8.4.- Control panel...
  • Page 6: Warnings

    DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian USER MANUAL Thanks for purchasing the Sd-300 Scandinavian dental equipment. This instruction manual contains information about the dental equipment, its configuration and maintenance. 0434 This symbol means the unit is certified under Directive 93/42/CEE (modified in accordance with 2007/47/EC) by the DNV Notes This symbol means CAUTION, PRECAUTION...
  • Page 7: Identification

    USER MANUAL DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian 3.- IDENTIFICATION The identification tag, containing technical information required for connection, can be found close to the base on the front of the chair, and is visible when the outer cover is removed (Fig. 1) SN: E (electronic unit)+number + S ExxxxS (Scandinavian) 4.- PRECAUTIONS...
  • Page 8: Clauses

    DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian USER MANUAL ENVIRONMENTAL PROTECTION All packaging materials are produced in respect for the environment and are fully recyclable: wooden pallets, cardboard, plastic bags and, bubble-wrap. Collecting used materials helps collection and recycling, and reduces waste material. Antoni Carles, S.A., is obliged to satisfy the objectives set by Community Directives 2002/95/ EC and 2002/96/EC.
  • Page 9: Technical Features

    USER MANUAL DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian 6.- TECHNICAL FEATURES 6.1.- Chair Ergonomic design for the patient. Dental chair with silent, automatic and highly reliable movement. Microprocessor-controlled movements: - Automatic chair return movement. - Automatic chair-to-spittoon movement.. - 3 programmable chair settings.
  • Page 10 DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian USER MANUAL Cannula support safety fitted as standard, with control pedal and chair safety features. Movement safety features in keypads and control pedal. Digital timer for basin (up to 3 minutes) and cup taps (up to 20 seconds). Cup filler and basin can be configured in joint or independent working selection.
  • Page 11: Transport And Dimensions

    USER MANUAL DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian 7.- TRANSPORT AND DIMENSIONS The dental unit is suitably packaged and protected. (Fig. 2) The package includes the dental unit secured to a pallet, with a further two cases containing the water unit and colibri forceps system.
  • Page 12: Composition

    DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian USER MANUAL 8.- SD300 SCANDINAVIAN COMPOSITION (Figure 3) Fig. 4. Circuit box Page 9 Fig. 5. Pedal Page 10 Fig. 6 Instrument tray Page 12 Fig. 7. Touch Expert control panel Page 14 Fig.
  • Page 13: Circuit Box

    USER MANUAL DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian 8.1.- Circuit box The circuit box contains all points for connecting the dental unit to the clinic’s power supply, as well as controls for adjusting the air and water supply. Regulator directional movement conforms to Standard UNE20128.
  • Page 14 DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian USER MANUAL 8.2.- Instrument control pedal (adjustable) It proportionally regulates the optional instruments on the instrument tray: micromotor, turbine and ultrasound. Regulation of the electric micro-motor’s turning speed and power of the turbine; with air only, or with air and water, using the lever-operated spray function.
  • Page 15: Instrument Tray

    USER MANUAL DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian Safety: To block any button-activated memory, press the control pedal. When activating the control pedal, all chair movements are blocked. This pedal does not allow any automatic memory to be saved nor does it accept it, so preventing unexpected movements, except for return-to-zero.
  • Page 16 DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian USER MANUAL TOP OF TRAY: A. Touch Expert control panel. B. Touch display surface. C. (Positions 1, 2, 3, 4, 5) Instrument poles (poles: 1, 2, 3, 4: installed upon request). D. Silicon instrument pad which can be sterilised in the autoclave. E.
  • Page 17: Control Panel

    USER MANUAL DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian 8.4.- Control panel MAIN CHARACTERISTICS: Consult the user manual provided for the Touch Expert Content control element Contents: Interface interpretation guide Main screen General fields: Names of the doctor and centre, date and time.
  • Page 18 DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian USER MANUAL CHAIR MOVEMENT FUNCTIONS - Safety for movements involved in activating instruments using the adjustment pedal. - Functions for activating movements parallel to the assistant keyboard (auxiliary) and adjustment pedal (there are no automatic movements on this, except for zeroing).
  • Page 19: Water Cabinet

    USER MANUAL DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian 8.5.- Water Cabinet As shown in the figure, the water unit has a turn of approx. 70° away from the arm support (H), facilitating essential movement for the surgical suction unit (D). The column arm (G) supports the dental light and instrument tray, and has a rotation of 220°.
  • Page 20: Cannula Support

    DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian USER MANUAL 8.6.- Cannula support (surgical suction, instruments, keypad) The tube support is mounted on an arm with a large capacity of movement to find the desided position. (Fig. 10) The surgical suction motor is activated by lifting any of the cannulas. Cannulas (A) can be sterilised, and are autoclave-safe;...
  • Page 21: Auxiliary Keypad

    USER MANUAL DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian 8.7.- Auxliary keypad The auxiliary keypad (or assistant’s keypad) is located on the cannula support. (Fig. 11) Fig. 11 F. Cup tap and spittoon tap. Timed, 1s-20 secs. It is possible to set up spittoon and cup tap to work together or separately. G.
  • Page 22: Headrest

    DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian USER MANUAL 8.8.- Headrest Thanks to its articulated movement, this head support allows the patient’s head to be positioned easily and ideally, according to the treatment required. It can be easily adjusted lengthways, as shown in Fig.
  • Page 23: Optional Instruments

    USER MANUAL DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian 8.10.- Optional instruments INSTRUMENT FITTED AS STANDARD 6F stainless steel syringe, always located to the left of all other instruments. The instrument has 6 functions (6F) which allow the water, air and spray functions to be operated in normal mode and connecting to the heat resistance.
  • Page 24 DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian USER MANUAL DISPLAY VIEW ERROR TYPE No micromotor(electric), or circuit broken, pole-1 No micromotor(electric), or circuit broken, pole-2 No micromotor(electric), or circuit broken, pole-3 Micromotor(electric) short-circuit or overload, pole-1 Micromotor(electric) short-circuit or overload, pole-2 Micromotor(electric) short-circuit or overload, pole-3 Light bulb missing or blown, instrument-pole-1 Light bulb missing or blown, instrument-pole-2...
  • Page 25: Safety Movements

    ANCAR thoroughly and systematically cleans your equipment, prior to its dispatch, with Magnolia Spray ECO-JET1 (Cattani Group). For cleaning and disinfecting the different parts of the system, please check: - Section “Cannula Support”.
  • Page 26: Technical Data

    DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian USER MANUAL 11.- TECHNICAL DATA Voltage 220-240 V~ Frequency 50 Hz Air pressure 5,5 Bar Water pressure 3 Bar Power 900 W Electrical protección type Type of operation Intermittent Maximum load (patient) 160 Kg.
  • Page 27 USER MANUAL DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian APPENDIX 1. CLEANING / DISINFECTING OF CATTANI GLASS, CANNULAS AND BASIN Before removing the water unit cover or extracting the vessel, disconnect the dental unit from the mains by turning off the mains. Use suitable protective gloves.
  • Page 28: Appendix 2. Metasys Amalgam Separator

    DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian USER MANUAL APPENDIX 2. METASYS AMALGAM SEPARATOR Before removing the water unit cover or extracting the vessel, disconnect the dental unit from the mains by turning off the general power switch. Fig. 16 Use suitable protective gloves.
  • Page 29: Appendix 3. Cleaning And Disinfecting Other Parts Of The Unit

    USER MANUAL DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian APPENDIX 3. CLEANING AND DISINFECTING OTHER PARTS OF THE UNIT. OTHER SAFETY PRECAUTIONS: UNIT, LAMP , ELECTRIC Cleaning and disinfecting Always disconnect the unit from the mains before carrying out any procedures. Do not wet or flood the unit with water.
  • Page 30 DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian USER MANUAL APPENDIX DURR VSA300 SUCTION SYSTEM. MAINTENANCE OF THE SUCTION SYSTEM TOGETHER WITH THE SPITTOON VALVE Before removing the cover of the water unit, disconnect the dental unit from the mains by deactivating the general power switch. Fig.
  • Page 31: Appendix 5. Minilight Syringe (Luzzani)

    USER MANUAL DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian APPENDIX 5. MINILIGHT SYRINGE (LUZZANI) GENERALIZATIONS The Minilight syringe is an instrument designed exclusively for dental use, its function being to introduce air and water (individually or simultaneously, at either room or body temperature) into the area of operation to keep it continuously clean and dry GENERAL CHARACTERISTICS The Minilight syringe has been designed using state of the art ergonomics to facilitate use...
  • Page 32 DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian USER MANUAL NORMAL USE - To introduce cold water into the area of operation, press the left button on the handle. - To introduce cold air into the area of operation, press the right button on the handle. - To introduce a cold air and water spray, press both buttons on the handle simultaneously.
  • Page 33: Appendix 6. Safety Observations

    USER MANUAL DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian APPENDIX 6. SAFETY OBSERVATIONS GENERAL PRECAUTIONS Read all necessary manuals. Keep all manuals – dental unit, chair, instruments, light and additional kits – in a safe place for future reference. Before starting up the unit, read all appendices in this manual.
  • Page 34 DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian USER MANUAL ELECTRICAL SAFETY MEASURES It is advisable not to use mobile phones in proximity of the dental unit. Observe regular hospital standards. In the event of a light overload, the built-in thermal protection in the 17 V ca line may have suspended operation of the unit.
  • Page 35 USER MANUAL DENTAL UNIT - QS4-451-01 / SD 300 Scandinavian...
  • Page 36 Antoni Carles, S.A. Volta dels Garrofers, 41-42 Pol. Ind. Els Garrofers 08340 - Vilassar del Mar Barcelona - (SPAIN) T. (34) 93 754 07 97 F. (34) 93 759 26 04 ancar@ancar-online.com www.ancar-online.com...