USER MANUAL DENTAL UNIT SD-150 - REV. 2 JULY CONTENTS PAGE 1. OVERVIEW 2. WARRANTY 3. IDENTIFICATION 4. PRECAUTIONS 5. CLAUSES 6. TECHNICAL FEATURES 6.1. Chair 6.2. Dental Unit 7. TRANSPORT AND DIMENSIONS 8. DESCRIPTION OF THE DENTAL UNIT 8.1. Connections box 8.2.
Defects, intended damage or modifications caused by third parties unrelated to Ancar. Non-compliance with the maintenance periods specified in the User Manual or resulting from inadequate repair or maintenance that has not been carried out by an official Ancar technical service.
Page 5
In no case shall ANCAR be required to replace the entire unit or to reimburse its purchase price. ANTONI CARLES, S.A.'s responsibility and accountability for any damages shall cease in the following cases, whereby the warranty shall be automatically cancelled: 1.
DENTAL UNIT SD-150 - REV.2 JULY 2020 USER MANUAL 3. IDENTIFICATION The identification label, which contains the technical information needed for connection, is folded. It can be seen from the rear of the seat and is also placed near the base, on the side of the chair.
Page 7
(ii) in all other countries, according to local regulations and recycling legislation. The dental equipment model Sd-150 complies with the essential requirements that apply to it as set forth in the Medical Device Directive 93/42/EEC and meets the design and construction requirements contained in standard EN60601-1-2 on the Safety of Medical Electrical Equipment and Electromagnetic Compatibility;...
DENTAL UNIT SD-150 - REV.2 JULY 2020 USER MANUAL ENVIRONMENTAL CONDITIONS When packaged for transport and storage, the equipment can withstand exposure up to 15 weeks to environmental conditions that do not exceed: a) ambient temperatures between -20˚C and +50˚C.
USER MANUAL DENTAL UNIT SD-150 - REV. 2 JULY 6. TECHNICAL FEATURES 6.1. Chair Ergonomic design for the patient. Chair movements powered by lifting gears that are silent and fully reliable. Movements controlled by a microprocessor: - Automatic return to zero movement (adjustable height).
DENTAL UNIT SD-150 - REV.2 JULY 2020 USER MANUAL 7. TRANSPORT AND DIMENSIONS The dental unit is delivered perfectly packaged and protected. One box contains the dental chair attached to a pallet and the other two boxes contain the hydric box and whip arm system, respectively. (Figure 2) It is essential that none of these boxes undergo impacts during transport and they must not fall on the ground under any circumstance.
DENTAL UNIT SD-150 - REV.2 JULY 2020 USER MANUAL 8.1. Connections box The connections box contains all the connections required to connect the dental equipment to the clinic's mains power supply and the air and water adjustment components. Valve movement direction complies with standard UNE 20128.
USER MANUAL DENTAL UNIT SD-150 - REV. 2 JULY 8.2. Electro-pneumatic pedal: USE AND INSTALLATION This pedal is used for proportional adjustment of the optional instruments in the instrument tray: micromotor, turbine. Ultrasound On/Off (Fig. 5). Regulates the rotation speed of the electrical micromotor, turbine power and pneumatic micromotor by moving the handle (B);...
DENTAL UNIT SD-150 - REV.2 JULY 2020 USER MANUAL 8.3. Instrument tray The instrument tray (Fig. 6) has an ergonomic design to facilitate the work of the user and an auxiliary instrument tray (A), slightly angled toward the main keypad (B) to optimise viewing, as well as a comfortable handle for movements (C).
Page 15
USER MANUAL DENTAL UNIT SD-150 - REV. 2 JULY The tray is joined to the equipment by a pantograph arm with a pneumatic brake to ensure a comfortable, effortless position at the front (handle side).
DENTAL UNIT SD-150 - REV.2 JULY 2020 USER MANUAL 8.4. Main keypad A. Activation of the dental operation lamp (ON / OFF). B. Tank and cuspidor water. Timer, 2-12 seconds. If the basin and water have been set to work at the same time, both will be activated.
USER MANUAL DENTAL UNIT SD-150 - REV. 2 JULY 8.5. Hydric unit The hydric unit rotates approximately 70° in relation to the support arm, increasing ergonomic comfort for daily work. The upper part contains the tank support (A) and basin (B), made of porcelain to provide a high degree of hygiene (Fig.
Page 18
DENTAL UNIT SD-150 - REV.2 JULY 2020 USER MANUAL The porcelain part of the basin can be easily removed by following the steps in Figure 11 and those in Fig.12 for the tap. Remove the tap first. (Fig. 12). To remove the tap, first pull up (“1”) on the tank tap (G), then remove (“2”) the tank support (H) to access the basin tap (J) and remove it (“3”).
USER MANUAL DENTAL UNIT SD-150 - REV. 2 JULY 8.6. Hose Support (surgical suction, instruments, auxiliary keypad) The hose support is mounted on an arm with a high capacity of movement to reach the desired position. The hose support is equipped with an auxiliary keypad (A), medium and full suction (B and C) and can be delivered with optional instruments, such as the 3F syringe (always placed in position E), a polymerising lamp or intra-oral camera (pos.
DENTAL UNIT SD-150 - REV.2 JULY 2020 USER MANUAL 8.7. Auxiliary keypad. The auxiliary keypad (or assistant keypad) is located on the hose support (Fig. 17). A. Tank and cuspidor water. Timer, 2s-12s. The cuspidor and basin can be set up to operate in combination or separately.
USER MANUAL DENTAL UNIT SD-150 - REV. 2 JULY 8.8. Headrest The headrest can easily be moved to the ideal position for the patient's head thanks to its articulation mechanism, depending on the treatment required. It is easy to adjust along its entire length, as shown in Figure 18.
A variety of dental hygiene specialists offer a wide range of cleaning products to achieve optimal results. ANCAR carries out systematic, comprehensive cleaning of its equipment before shipping with the disinfecting agent ECO-JET1 Spray by Magnolia (Cattani Group). To clean and disinfect the various parts of the equipment, please refer to: - The “Hose Support”...
USER MANUAL DENTAL UNIT SD-150 - REV. 2 JULY 11. TECHNICAL DATA Voltage 230V Frequency 50 Hz Air pressure 5 Bar Water pressure 2.8 Bar Power 900 W Electrical protection class Type of operation Intermittent Maximum load (patient) 160 Kg.
DENTAL UNIT SD-150 - REV.2 JULY 2020 USER MANUAL On request, the hydric unit can be supplied with a drain by gravity or continuous suction system and with the option of an amalgam separator for wet or dry systems. See the annexes of this manual to follow each manufacturer's specific instructions.
USER MANUAL DENTAL UNIT SD-150 - REV. 2 JULY ANNEX VS300 SUCTION VALVES (DÜRR DENTAL). MAINTENANCE OF THE SUCTION SYSTEM IN COMBINATION WITH THE CUSPIDOR VALVE Before opening the door of the hydric box or removing the tank, disconnect the dental unit from the mains by switching the main switch off.
DENTAL UNIT SD-150 - REV.2 JULY 2020 USER MANUAL ANNEX 3. TYP1 SUCTION SYSTEM (METASYS) Before opening the door of the hydric box or removing the tank, disconnect the dental unit from the mains by switching the main switch off.
USER MANUAL DENTAL UNIT SD-150 - REV. 2 JULY ANNEX 4. ECO SUCTION SYSTEM (METASYS) Before opening the door of the hydric box or removing the tank, disconnect the dental unit from the mains by switching the main switch off.
DENTAL UNIT SD-150 - REV.2 JULY 2020 USER MANUAL ANNEX 5. SELF WATER SYSTEM KIT Access the bottle (1-litre capacity) from the outer part of the hydric unit by removing the cover (A) outwards ("1"). This is an independent mains water system; the air regulator is normally sealed.
USER MANUAL DENTAL UNIT SD-150 - REV. 2 JULY ANNEX 6 CAS 1 COMBI-SEPARATOR (DÜRR DENTAL) The control panel is located on the inside (A). The tank (B) should be cleaned. To remove it, unscrew (“1”), but before extracting it (“2”), note that there is a part that measures the level of the amalgam elements contained that protrudes from the bottom.
DENTAL UNIT SD-150 - REV.2 JULY 2020 USER MANUAL ANNEX 8. WATER DECONTAMINATION: WEK (METASYS) The bag of product (B) is inside the mixing system. The design allows the water to arrive after a distance of free fall, thanks to which the separation between the WEK system and the supply of water from the mains guarantees compliance with the DVGW standard.
USER MANUAL DENTAL UNIT SD-150 - REV. 2 JULY ANNEX 9. H1 HYGIENE SYSTEM (METASYS) Do not use any additional cleaning or disinfecting product for the suction installation in treatment areas where an H1 hygiene system has been installed. There is a risk of harmful chemical interactions due to the combination of different chemical products.
DENTAL UNIT SD-150 - REV.2 JULY 2020 USER MANUAL AUTOMATIC CLEANING PROGRAM First Wash Duration: approximately 2 minutes. Led F will flash slowly. Start-up: press on the main switch of the treatment unit. Function: preparation of later treatments with a higher dose of disinfectants.
Page 33
USER MANUAL DENTAL UNIT SD-150 - REV. 2 JULY MANUFACTURING BATCH Each product is identified with an internal number placed at the end that identifies the exact production batch it belongs to. This number indicates the period of manufacture associated with the control card.
Page 34
DENTAL UNIT SD-150 - REV.2 JULY 2020 USER MANUAL CLEANING AND STERILISATION After each intervention, the syringe should be cleaned and sterilised to guarantee maximum hygiene. Do so by taking the following steps: 135° - Remove the tip (unscrew the nose-tip terminal) and/or the entire grip (by pressing the button located at the bottom of the same and pressing upwards).
USER MANUAL DENTAL UNIT SD-150 - REV. 2 JULY ANNEX 11. OPTIONAL COMPONENTS AUXILIARY AIR AND WATER OUTLETS / USB INPUT In the hydric unit it is optional to have a rapid water (C) and air (D) outlet, as well as a USB input (A) for end-to-end communication, the other end of which will pass through the dental unit.
DENTAL UNIT SD-150 - REV.2 JULY 2020 USER MANUAL ANNEX 12. CLEANING AND DISINFECTION OF OTHER EQUIPMENT PARTS Cleaning and disinfection Carry out the operations with the device disconnected from the mains. Do not flood or wet the equipment with water.
Page 37
USER MANUAL DENTAL UNIT SD-150 - REV. 2 JULY ANCAR recommends the use of the reloadable BODE X-WIPES dispenser for professional cleaning and disinfection. Indicated for all BODE surface disinfectants in concentrations that act for 1 hour or less.
DENTAL UNIT SD-150 - REV.2 JULY 2020 USER MANUAL ANNEX 13. SAFETY OBSERVATIONS GENERAL PRECAUTIONS Refer to all the necessary manuals. Keep all the manuals - dental unit, chair, instruments, lamp and additional kits - in a safe place for future reference.
Page 39
USER MANUAL DENTAL UNIT SD-150 - REV. 2 JULY Risk of intervention with autonomous devices (e.g., motor for implants). Disconnect the power of the dental equipment to prevent indirect movements caused by malfunctions and/or unexpected activation of movement controls. This equipment complies with standard EMC (EN 60601-1-2:2007 + AC:2010).
DENTAL UNIT SD-150 - REV.2 JULY 2020 USER MANUAL Electromagnetic immunity Test Category Comments Electrostatic discharge Discharge into the air: 2, 4, 8 kV Improved performance with wood and ceramic materials. Contact discharge: 2, 4, 6 kV With synthetic materials, humidity >...
Page 41
USER MANUAL DENTAL UNIT SD-150 - REV. 2 JULY...
Page 42
08340 Vilassar de Mar (Barcelona-SPAIN) Tel. (34) 93 754 07 97 Fax (34) 93 759 26 04 www.ancar-online.com / e-mail: ancar@ancar-online.com Subject to design and/or specifications changes DENTAL UNIT SD-150 - REV. 2 JULY 2020 without prior notice. Made in Europe. 0197...