Ancar SD 150 User Manual
Hide thumbs Also See for SD 150:

Advertisement

Quick Links

SD 150
USER MANUAL

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SD 150 and is the answer not in the manual?

Questions and answers

Summary of Contents for Ancar SD 150

  • Page 1 SD 150 USER MANUAL...
  • Page 3 Antoni Carles, S.A. Volta dels Garrofers, 41-42 T. (34) 93 754 07 97 Pol. Ind. Els Garrofers F. (34) 93 759 26 04 08340-Vilassar de Mar calidad@ancar-online.com ISO 9001 ISO 13485 (Barcelona-SPAIN) www.ancar-dental.com 109037-AQ-IBE 112630-AQ-IBE EC Declaration of Conformity (Directive 93/42/EEC) Manufacturer’s Name...
  • Page 5: Table Of Contents

    USER MANUAL DENTAL UNIT - QS4-366-02 / SD 150 TABLE OF CONTENTS PAGE 1.- PRECAUTIONS 2.- GUARANTEE 3.- IDENTIFICATION 4.- PRECAUTIONS 5.- CLAUSES 6.- TECHNICAL FEATURES 6.1.- Chair 6.2.- Dental Unit 7.- DIMENSIONS AND TRANSPORTATION 8.- DESCRIPTION 8.1.- Circuit box 8.2.- Pneumatic pedal...
  • Page 6: Precautions

    DENTAL UNIT - QS4-366-02 / SD 150 USER MANUAL Thank you for purchasing the SD-150 dental unit. This instruction manual contains information on the dental unit, including its configuration and maintenance. 0434 This symbol means the unit is certified under Directive 93/42/EEC by...
  • Page 7: Identification

    USER MANUAL DENTAL UNIT - QS4-366-02 / SD 150 3.- IDENTIFICATION The identification tag, containing technical information required for connection, can be found close to the base on the side of the chair. It is visible when the outer cover is removed (Fig. 1).
  • Page 8: Clauses

    DENTAL UNIT - QS4-366-02 / SD 150 USER MANUAL ENVIRONMENTAL PROTECTION All packaging materials are produced in respect for the environment and are fully recyclable: wooden pallets, cardboard, plastic bags and bubble-wrap. Collecting used materials helps collection and recycling and reduces waste material.
  • Page 9: Technical Features

    USER MANUAL DENTAL UNIT - QS4-366-02 / SD 150 6.- TECHNICAL FEATURES 6.1.- Chair Ergonomic design for the patient. Dental chair with silent, automatic and highly reliable movement. Microprocessor-controlled movements: - automatic chair return movement. - automatic chair-to-spittoon movement. - 3 programmable chair settings.
  • Page 10: Dimensions And Transportation

    DENTAL UNIT - QS4-366-02 / SD 150 USER MANUAL 7.- DIMENSIONS AND TRANSPORTATION The dental unit is suitably packaged and protected. The package includes the dental unit secured to a pallet, with a further two cases containing the water unit and colibri forceps system (Fig. 2).
  • Page 11: Description

    USER MANUAL DENTAL UNIT - QS4-366-02 / SD 150 8.- SD150 DENTAL UNIT DESCRIPTION (Fig. 3) Fig. 4 Circuit box Page 8 Fig. 5 Pneumatic pedal Page 9 Fig. 6 Instrument tray Page 11 Fig. 7 Main control panel Page 12 Fig.
  • Page 12: Circuit Box

    DENTAL UNIT - QS4-366-02 / SD 150 USER MANUAL 8.1.- Circuit box The circuit box contains all points for connecting the dental unit to the clinic’s power supply, as well as controls for adjusting the air and water supply. Regulator directional movement conforms to Standard UNE 20128.
  • Page 13: Pneumatic Pedal

    USER MANUAL DENTAL UNIT - QS4-366-02 / SD 150 8.2.- Pneumatic pedal: INSTALLATION AND USE Pedal for adjusting the optional instruments on the instrument tray: Micromotor, turbine. Ultrasound on/off switch. Control the electric micromotor rotation speed, turbine power and pneumatic micromotor by activating the trigger (B);...
  • Page 14: Instrument Tray

    DENTAL UNIT - QS4-366-02 / SD 150 USER MANUAL 8.3.- Instrument tray The instrument tray (Fig. 6), ergonomically designed to facilitate the work of the user, is also fitted with an auxiliary instrument tray (A), a slight inclination of the main keypad (B) to provide easy viewing, and a comfortable chair positioning handle (C).
  • Page 15 USER MANUAL DENTAL UNIT - QS4-366-02 / SD 150 The tray is attached to the unit by means of a pantographic arm, with a pneumatic brake for comfortable, effortless front positioning (handle side). Fig. 6...
  • Page 16: Main Keypad

    DENTAL UNIT - QS4-366-02 / SD 150 USER MANUAL 8.4.- Main keypad: (Fig. 7) 1. Dental light switch (ON/OFF) 2. Water cup and spittoon taps. Timed, 2-12 seconds. Spittoon and cup taps can be set-up to work simultaneously or independently.
  • Page 17: Water Cabinet

    USER MANUAL DENTAL UNIT - QS4-366-02 / SD 150 8.5.- Water Cabinet As seen in Fig. 8, the water unit has a rotation of 75° approx. towards the arm support (H), facilitating the ergonomics of daily use. The column arm (G) supports the dental light and instrument tray and has a rotation of 120°.
  • Page 18: Cannula Support

    DENTAL UNIT - QS4-366-02 / SD 150 USER MANUAL 8.6.- Cannula support (surgical suction, instruments, keypad) The tube support is mounted on an arm with a large capacity of movement to find the desired position (Fig. 9). The surgical suction motor is activated by lifting any of the cannulas. Cannulas (A) and cannula- holders (B) can be sterilised and are autoclave-safe.
  • Page 19: Auxiliary Keypad

    USER MANUAL DENTAL UNIT - QS4-366-02 / SD 150 8.7.- Auxiliary keypad. The auxiliary keypad (or assistant’s keypad) is located on the cannula support (Fig. 10). Fig. 10 F. Cup tap and spittoon tap. Timed, 2s-12s. Spittoon and cup taps can be set-up to work simultaneously or independently.
  • Page 20: Headrest

    DENTAL UNIT - QS4-366-02 / SD 150 USER MANUAL 8.8.- Headrest Thanks to its articulated movement, this head support allows the patient’s head to be positioned easily and ideally, according to the treatment required. It can be easily adjusted lengthways, as shown in Fig.
  • Page 21: Safety Movements

    Before being shipped, all ANCAR equipment is thoroughly cleaned using ECO-JET1 Cattani Magolia disinfectant spray. For cleaning and disinfecting the different parts of the system, please check: - Section “Cannula Support”.
  • Page 22: Technical Data

    DENTAL UNIT - QS4-366-02 / SD 150 USER MANUAL 11.- TECHNICAL DATA Voltage 220-240 V~ Frequency 50 Hz Air pressure 5.5 Bar Water pressure 3 Bar Power 900 W Electrical protection type Operation type Intermittent Maximum load (patient) 160 Kg.
  • Page 23: Appendix 1. Cleaning And Disinfecting Of Cattani Glass, Cannulas And Basin

    USER MANUAL DENTAL UNIT - QS4-366-02 / SD 150 APPENDIX 1. CLEANING AND DISINFECTING OF CATTANI GLASS, CANNULAS AND BASIN Before removing the water unit cover or taking out the vessel, disconnect the dental unit from the mains by turning off the general power switch. Use suitable protective gloves.
  • Page 24: Appendix 2. Metasys Amalgam Separator

    DENTAL UNIT - QS4-366-02 / SD 150 USER MANUAL APPENDIX 2. METASYS AMALGAM SEPARATOR Before removing the water unit cover or taking out the vessel, disconnect the dental unit from the mains by turning off the general power switch. Use suitable protective gloves.
  • Page 25: Appendix 3. Cleaning And Disinfecting Other Parts Of The Unit

    USER MANUAL DENTAL UNIT - QS4-366-02 / SD 150 APPENDIX 3. CLEANING AND DISINFECTING OTHER PARTS OF THE UNIT. OTHER SAFETY PRECAUTIONS: UNIT, LIGHT, ELECTRIC Cleaning and disinfecting Always disconnect the unit from the mains before carrying out any procedures.
  • Page 26: Appendix 4. Durr Vsa300 Suction System

    DENTAL UNIT - QS4-366-02 / SD 150 USER MANUAL APPENDIX DURR VSA300 SUCTION SYSTEM. MAINTENANCE OF THE SUCTION SYSTEM TOGETHER WITH THE SPITTOON VALVE Before removing the cover of the water unit, disconnect the dental unit from the mains by deactivating the general power switch.
  • Page 27: Appendix 5. Minilight Syringe (Luzzani)

    USER MANUAL DENTAL UNIT - QS4-366-02 / SD 150 APPENDIX 5. MINILIGHT SYRINGE (LUZZANI) GENERAL The Minilight syringe is an instrument designed exclusively for dental use, its function being to introduce air and water (individually or simultaneously, at either room or body temperature) into the area of operation to keep it continuously clean and dry.
  • Page 28 DENTAL UNIT - QS4-366-02 / SD 150 USER MANUAL INSTALLATION The unit may only be connected by an Antoni Carles-authorised technician. NORMAL USE - To introduce cold water into the area of operation, press the left button on the handle.
  • Page 29: Appendix 6. Safety Observations

    USER MANUAL DENTAL UNIT - QS4-366-02 / SD 150 APPENDIX 6. SAFETY OBSERVATIONS GENERAL PRECAUTIONS Read all necessary manuals. Keep all manuals – dental unit, chair, instruments, light and additional kits – in a safe place for future reference. Before switching on the unit, read all appendices in this manual.
  • Page 30 DENTAL UNIT - QS4-366-02 / SD 150 USER MANUAL ELECTRICAL SAFETY MEASURES It is advisable not to use mobile phones in proximity of the dental unit. Observe regular hospital standards. In the event of a light overload, the built-in thermal protection in the 17VCA line may have suspended operation of the unit.
  • Page 31 Antoni Carles, S.A. Volta dels Garrofers, 41-42 Pol. Ind. Els Garrofers 08340 - Vilassar del Mar Barcelona - (SPAIN) T. (34) 93 754 07 97 F. (34) 93 759 26 04 ancar@ancar-online.com www.ancar-online.com...

This manual is also suitable for:

34991

Table of Contents