Controls and Indicators

Controls

Symbol
Description
ON/OFF button - Press once to switch the Pump on . Press and hold down for approximately 3 seconds to switch
the Pump off .
a
Note: Logs are maintained for power down events including when the Pump is powered down or
d
RUN/HOLD button - Press to start the infusion or to put the infusion on hold .
CLEAR/MUTE button - Press to silence alarm or advisories for 1 minute . The audio from the alarm or advisories
m
will resound after this time . Resets numeric values to zero .
Note: To re-enable the alarm audio press any other button or key to unmute .
b
PRIME/BOLUS button - Primes the Infusion set . Administers bolus during the infusion .
PRIMARY/SECONDARY button - Switches the Pump between Primary and Secondary infusion modes . (If
l
enabled) .
ENTER button - Scrolls between rate, time, VTBI and total volume infused (VI) . Enters values for selected infusion
e
/ configuration parameters . Confirms the rate during an infusion titration .
CHEVRON keys - Increases or decreases the infusion rate, TIME limit and VTBI . Press and hold to increase the
hf
selection speed . Used to adjust user selectable options

Indicators: (when illuminated)

Symbol
Description
AC POWER indicator - When illuminated the Pump is connected to an AC power supply .
S
RATE
The Pump is displaying the infusion rate in millilitres per hour (ml/h) .
VTBI
The Pump is displaying the Volume To Be Infused (VTBI) in millilitres (ml) .
VI
The Pump is displaying the Volume Infused (VI) in millilitres (ml) .
TIME
The Pump is displaying the infusion time in hours : mins .
MICRO
The Pump is operating in the Micro Mode . When not illuminated the Pump is in the Standard Mode .
SEC
The Pump is operating in the Secondary Mode . When not illuminated the Pump is in the Primary Mode .
(Millilitres / hour) When ml is illuminated the Pump displays the rate, VTBI or VI . When the h is illuminated the
ml/h
Pump displays the rate or infusion time .
ALARM Indicator - When flashing indicates the Pump is in an alarm condition .
MUTE Indicator - When flashing indicates the Pump audio alarm is muted .
unexpected power loss .
BDDF00160 Issue 1
Alaris™ GW Volumetric Pumps
Controls and Indicators
6/40