Trumpet and Flow Rate Curves

In this Pump, as with all infusion systems, the action of the pumping mechanism and variations cause short-term fluctuations in rate
accuracy .
The following curves show typical performance of the system in two ways: 1) the accuracy of fluid delivery over various time periods is
measured (trumpet curves), and 2) the delay in onset of fluid flow when infusion commences (start-up curves) .
Trumpet curves are named for their characteristic shape . They display discrete data averaged over particular time periods or 'observation
windows' , not continuous data versus operating time . Over long observation windows, short term fluctuations have little effect on
accuracy as represented by the flat part of the curve . As the observation window is reduced, short term fluctuations have greater effects
as represented by the mouth of the trumpet .
Knowledge of system accuracy over various observation windows may be of interest when certain drugs are being administered . Short
term fluctuations in rate accuracy may have clinical impact depending on the half-life of the drug being infused and the degree of inter
vascular integration, the clinical effect cannot be determined from the trumpet curves alone .
The start-up curves represent continuous flow versus operating time for two hours from the start of the infusion . They exhibit the delay
in onset of delivery due to mechanical compliance and provide a visual representation of uniformity . Trumpet curves are derived from
the second hour of this data . Tests performed per IEC60601-2-24 standard .
The trumpet and flow rate curves are representative of the typical performance of the Alaris GW 800 Volumetric
w
Pump when used in combination with a typical dedicated infusion set. The user is responsible for assessing the
performance of the Alaris GW 800 Volumetric Pump when used in combination with a non-dedicated infusion set.
Start-up Graph at 1 .0ml/h (Initial Period)
Start-up Graph at 1.0 ml/h (Initial Period)
273-001 Infusion Set
2.0
1.8
1.6
1.4
1.2
1.0
0.8
0.6
0.4
0.2
0.0
-0.2
0
10
20
30
40
Time (minutes)
Trumpet Graph at 1 .0ml/h (Initial Period)
273-001 Infusion Set
Trumpet Graph at 1.0 ml/h (Initial Period)
60
50
40
30
20
10
0
-10
-20
-30
-40
-50
-60
0
5
10
15
Observation Window (Minutes)
Maximum
Minimum
Rate Error
Rate Error
Trumpet Graph at 1 .0ml/h (After 24 Hours)
273-001 Infusion Set
Trumpet Graph at 1.0 ml/h (After 24 Hours)
80
70
60
50
40
30
20
10
0
-10
-20
-30
-40
-50
-60
-70
-80
0
5
10
15
Observation Window (Minutes)
Maximum
Minimum
Rate Error
Rate Error
Note: The typical flow rate and trumpet curves - Infusion set 273 001
50
60
70
80
90
100
110
120
Time (minutes)
20
25
30
35
Overall Mean Error
= +2 .5%
20
25
30
35
Overall Mean Error
= +4 .2%
BDDF00160 Issue 1
Alaris™ GW Volumetric Pumps
Start-up Graph at 25 .0ml/h (Initial Period)
Start-up Graph at 25.0 ml/h (Initial Period)
273-001 Infusion Set
50.0
45.0
40.0
35.0
30.0
25.0
20.0
15.0
10.0
5.0
0.0
0
10
20
30
40
50
60
Time (minutes)
Time (minutes)
Trumpet Graph at 25 .0ml/h (Initial Period)
273-001 Infusion Set
Trumpet Graph at 25.0 ml/h (Initial Period)
15
10
5
0
-5
-10
-15
0
5
10
15
Observation Window (Minutes)
Maximum
Minimum
Rate Error
Rate Error
Trumpet Graph at 25 .0ml/h (After 24 Hours)
273-001 Infusion Set
Trumpet Graph at 25.0 ml/h (After 24 Hours)
15
10
5
0
-5
-10
-15
0
5
10
15
Observation Window (minutes)
Observation Window (Minutes)
Maximum
Minimum
Rate Error
Rate Error
37/40
Trumpet and Flow Rate Curves
70
80
90
100
110
120
20
25
30
35
Overall Mean Error
= +1 .1%
20
25
30
35
Overall Mean Error
= +1 .1%