Configurable Options Sheet - BD Alaris GW 800 Series Directions For Use Manual

Volumetric pumps
Hide thumbs Also See for Alaris GW 800 Series:
Table of Contents

Advertisement

Configurable Options Sheet

The default settings are configurable as displayed in brackets in the table below. Each of the configurable options
w
has a code which must only be altered by Qualified Service Personnel with reference to the technical service manual
(TSM) for this product (Technical Service Manual reference: BDTM00006).
Description
Enable VTBI / time infusions
Maximum priming volume
Clear infusion parameters to zero on
power - up
Maximum VTBI in MICRO Mode
Bolus rate
Maximum bolus volume
Keep vein open rate
Air in line alarm volume - single bubble
Enable secondary infusion capability
Default occlusion pressure on power - up (Lo (250mmHg),Nor (350mmHg), HI (500mmHg)) HI
Alarm volume level
Enable Micro mode
Maximum infusion rate
Enable ASCII mode for communications
Enable Odd parity for communications
Set Pump address for communications
Flow Sensor Connection Mode**
Set - up of current time and date
Language selection
IrDA Communications Selection
Nurse Call Activation High Enabled
Drops per ml of fluid
Enable Automatic Set Detection and Anti-
bolus Function***
Silent Mode
User select mode options
Pressure limit Enabled
Alarm volume Enabled
Timed infusions Enabled
Micro infusions Enabled
Flow sensor sensitivity level
*EnGL - English, FrAn - French, dEut - German, ItAL - Italian, ESPA - Spanish, SE - Swedish, nEd - Dutch.
** If infusion sets without anti-siphon valves are being used, it is recommended to change the flow sensor connection mode setting to On. With
On selected, the Alaris GW 800 Volumetric Pump will not operate unless a flow sensor is connected.
*** When using dedicated BD Infusion sets (refer to Compatible Dedicated Infusion Sets section) ensure that the set detection mode is On (Refer
to Automatic Set Detection section for more details). Align all Alaris GW 800 Volumetric Pumps in your hospital to the same set detection mode
(ie. all On or all OFF) to avoid potential confusion amongst users as to the mode in which the Pump is operating under.
Serial Number
Configured by
Approved by
Range
(On / OFF)
(OFF, 1 to 40 ml)
(On / OFF)
(0 .1 to 999 ml)
(1 to 999ml/h)
(OFF, 1 to 99ml)
(OFF, 1 .0 to 5 .0 ml/h)
(50,100, 250, 500µL)
(On / OFF)
(1 - 7)
(On / OFF)
(1 - 999 ml/h)
(On / OFF)
(On / OFF)
(1 - 250)
(AUTO/On)
(00:00 to 23:59)
(01/01/00 to 31/12/99)
(EnGL, FrAn, dEut, ItAL, ESPA, SE, nEd)*
(On / OFF)
(On / OFF)
(1 to 200)
(On / OFF)
(On / OFF)
(On / OFF)
(On / OFF)
(On / OFF)
(On / OFF)
(Nor, Hi)
BDDF00160 Issue 1
Alaris™ GW Volumetric Pumps
Default
OFF
40ml
OFF
999ml
400ml/h
5ml
5 .0ml/h
100µl
OFF
4
OFF
999ml/h
OFF
OFF
1
On
N/A
EnGL
On
On
20
OFF
OFF
OFF
OFF
OFF
OFF
Nor
Software Version
Date
Date
23/40
Configurable Options Sheet
Setting