Automatic Set Detection - BD Alaris GW 800 Series Directions For Use Manual

Volumetric pumps
Hide thumbs Also See for Alaris GW 800 Series:
Table of Contents

Advertisement

Alaris™ GW Volumetric Pumps
Getting Started

Automatic Set Detection

When using a compatible dedicated BD infusion set (refer to 'Compatible Dedicated Infusion Sets' section), it is recommended that the
set detection mode is enabled on the Alaris GW 800 Volumetric Pump . This mode can be enabled in Technician Mode on the Pump .
Align all Alaris GW 800 Volumetric Pumps within a particular hospital or clinical environment to the same automatic set detection mode
(ie . All On or all OFF) . This will help to avoid potential confusion amongst users as to the mode in which the Pump is operating under and
hence to ensure that the appropriate recommendations for flow sensor usage are followed .
When the automatic set detection is set to On, the Pump automatically tests that a compatible BD Infusion set has been loaded correctly
(refer to 'Compatible Dedicated Infusion Sets' section) . The test will occur at the start of the first infusion after the Pump is switched on or
after the door has been opened; the Pump will run in reverse for 10 seconds and then forward for 10 seconds, the test takes a maximum
of 20 seconds to complete . During this operation the clinician may observe a blood return that will be more evident if using a small
catheter .
If the Pump fails to detect a correct BD Infusion set or detects a possible set misload, then the Pump will alarm and display
Q
refer to 'Alarms' section of this DFU .
Please contact your local BD representative if further information or support is required regarding the Automatic Set Detection
operation or the application of this Pump in specific clinical settings, e .g . neonatal .
BDDF00160 Issue 1
16/40

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents