User Selectable Options

The pump must be on HOLD. Press

Check the Battery Status

Note:
Only available when Pump is operating on battery power.
Press and hold
for 2 seconds until Battery Status is displayed:
e
,
.
e
Press
button to go to next option.
Selecting the Occlusion Pressure Level (if Enabled)
PrES option is displayed:
s
Selecting the Alarm Volume (if Enabled)
tonE option is displayed:
t
Selecting VTBI / Time Infusion (if Enabled)
CLoc option is displayed:
q
Selecting Micro Mode (if Enabled)
0.0 option is displayed:
r
.
d
• Battery has greater than 30 minutes run time.
• Battery is low and has approximately 30 minutes or less
run time.
• Change occlusion pressure level to high (HI), normal (nor)
or low (lo) using
h
• Press
to confirm and go to next option.
e
• Change alarm volume between 1 and 7* using
• Press
to confirm and go to next option.
e
* 1=Low, 7=High
• Select On or OFF using
infusion** setting on or off.
• Press
to confirm and go to next option.
e
** Ensure TIME indicator is lit, if set to On.
• Select On or OFF using
on or off.
• Press
to confirm.
e
*** Ensure MICRO indicator is lit, if set to On.
0000CF04392 Iss. 1 7/12
/
.
f
/
keys to turn the VTBI/time
h
f
/
keys to turn micro mode***