SWR M-Sens 2 Manual page 22

Online-moisture meter for solids
Table of Contents

Advertisement

22
5.5 Alarm value
Choose the output value for the alarm (sensor error
or an internal alarm); at the same time the relay 3
drops down. Range of 0 ... 22 mA (Std 3 mA)
With C you set the value to 0.0. With the num-
bers you input the beginning value and with  the input is accepted and you quit
the menu level.
5.6 Filter time
Choose the filter time for the current output in the
range of 0,1 ... 999,9 s (Standard 1 s)
With C you set the value to 0.0. With the num-
bers you input the beginning value and with 
the input is accepted and you quit the menu level.
5.7 Calibration: 4mA
Set the minimal current output. Fit to the external
measurement system (at differing display).
Adapt the current output to 4 mA with the keys < <
and >> quickly and with the keys < und > slowly.
With  the input is accepted and you quit the menu level, with C you quit the
menu point without any change.
5.8 Calibration: 20mA
Set the maximal current output. Fit to the external
measurement system (at differing display).
Adapt the current output to 4 mA with the keys < <
and >> quickly and with the keys < und > slowly.
With  the input is accepted and you quit the menu level, with C you quit the
menu point without any change.
6. Digital inputs
It is only necessary for stopping and starting the
measurement via an external control line (for wir-
ing see 8.1).
With  and  you make a choice according to the display. With C you quit the
menu point without any change and with  the input is accepted and you arrive 1
menu point lower to make your settings.
6.1 Digital input 1
With  and  you make a choice according to
the accentuation. With C you quit the menu point
without any change and with  the input is ac-
cepted and you arrive 1 menu point lower to make
your settings.
Anal. output
7
8
Alarm value
4
5
3.0
mA
1
2
C
0
Anal. output
7
8
Filter time
4
5
3.0
s
1
2
C
0
Anal. output
Calibration 4.0mA
< < <
> >>
Anal. output
Calibration 20.0mA
< < <
> >>
6.Digital inputs
6.1.Digital input 1
6.2.Digital input 2
6.1. Dig. input1
6.1.1.Funct.
none
6.1.2.Direct. direct
6.1.3.Filter
1.0s
9
6
3
9
6
3
C
C
C
C

Advertisement