1.4.10 Calibration point N
Choose the measured value in physical unities – beginning and end of
measurement range.
1.4.11 Measured value
Input value is collected and assigned to the shown measured value.
For 2.1 to 2.4 / 3.1 to 3.4 / 4.1 to 4.4 : analogue to point 1
5. Analogue output
With  and  you make a choice according to
the accentuation. With C you quit the menu point
without any change and with  the input is ac-
cepted and you arrive 1 menu point lower to make
your settings.
5.1 Beginning of range
Choose the output in the range of 0 ... 22 mA
(Standard 4 mA).
With C you set the value to 0.0. With the num-
bers you input the beginning value and with 
the input is accepted and you quit the menu level.
5.2 End of range
Choose the output in the range of 0 ... 22 mA
(Standard 20 mA).
With C you set the value to 0.0. With the num-
bers you input the beginning value and with 
the input is accepted and you quit the menu level.
5.3 MIN limit
Choose the minimal output value in the range of
0 ... 22 mA (Standard 3 mA).
With C you set the value to 0.0. With the num-
bers you input the beginning value and with 
the input is accepted and you quit the menu level.
5.4 MAX limit
Choose the minimal output value in the range of
0 ... 22 mA (Standard 20 mA).
With C you set the value to 0.0. With the num-
bers you input the beginning value and with 
the input is accepted and you quit the menu level.
5.Analog. outupt
5.1.Beginning
4.0mA
5.2.End
20.0mA
5.3.MIN
0.3mA
5.4.MAX
21.0mA
Anal. output
7
8
Range beg.
4
5
4.0
mA
1
2
C
0
Anal. output
7
8
Range end
4
5
20.0
mA
1
2
C
0
Anal. output
7
8
MIN limit
4
5
3.0
mA
1
2
C
0
Anal. output
7
8
MAX limit
4
5
20.0
mA
1
2
C
0
C
9
6
3
9
6
3
9
6
3
9
6
3