This Field Service Bulletin provides information on the following per ECO# 69435:
• Introduces revision "D" release of the VersaPulse® PowerSuite™ software, which corrects
the following bugs:
* Removes the word TREATMENT from the screen which was displayed after the sys-
* Adds a 2 J @10 Hz Holmium operating parameter to the 20/60W model.
* Disables the "Auto Shut-off" feature in the burn-in software in all field units manu-
• Orders an inspection of the rear panel cable harness and fill tube routing near and around
the water pump motor on all field units manufactured before July 2001 on a next call basis.
PARTS REQUIRED
Qty. 1
P/N 0638-555-01
Qty. 2
Electrical Tape
RETROFIT INSTRUCTIONS
Perform this retrofit on a next call basis.
1. Disconnect the system power cord from the AC mains.
2. Gain access to the rear panel.
Open the front door and remove the top and side covers.
Remove the rear panel assembly mounting hardware from the chassis and carefully lean the top of the
rear panel assembly away from the chassis.
2 JULY 2001
tem had dropped to STANDBY as a result of a system fault.
factured before February 2001.