Do you have a question about the SELECTA and is the answer not in the manual?
Questions and answers
JANE
March 31, 2025
WHAT IS SAVER MODE ON MY LASER
1 comments:
Mr. Anderson
May 14, 2025
Saver Mode, referred to as CaseSaver™ Mode, is a feature on the LUMENIS SELECTA laser that allows continued operation if one of the laser wavelengths becomes inoperable. If the unit has more than one wavelength, the remaining operational wavelengths continue to function, enabling treatment to proceed until the unit is serviced.
Page 1
Selecta ® Ophthalmic Laser System Operator Manual ®...
Page 2
Lumenis, the Lumenis Logo, and Novus are registered trademarks of Lumenis, Inc. InSight, SureSpot, Varia, and Spectra are trademarks of Lumenis, Inc. Manufactured in Italy by C.S.O. S.r.l. for:...
Introduction 9 Introduction The Selecta ophthalmic laser systems are classified as Class IIIb lasers by the Center for Devices and Radiological Health of the Food and Drug Administration and as Class 3b by the International Standard IEC 60825. Users must take precautions to prevent exposure of laser energy to the eyes and skin from either direct or diffusely reflected laser beams, except as a therapeutic application.
(usually within the controlled area). For additional information, refer to ANSI Z136.1-2000, ANSI Z136.3-2005, or International Standard IEC 60825-1: 2001. The following formula was used to calculate the worst case NOHD for Lumenis Selecta lasers and compatible delivery systems: Φ Pf - a NOHD = θ...
Page 11
NOHD = Laser system Using this approach we derive the following values: Laser System Selecta Duet 1064 nm laser 0.262 0.03 J 0.0007 cm 9.3 cm 1.67 (10 ) J/cm Selecta Duet 532 nm laser 0.016 0.002 J...
Page 12
The barrier should be made of material that can withstand the power of the treatment beam for the maximum exposure time, relative to the configuration of the controlled area and the treatment parameters for the specific medical application. Selecta PBL200051, Rev J...
WARNING - Never look directly into any optical lens, except for therapeutic purposes, nor any optical fiber, probe, or laser system aperture while the laser is energized. Severe eye damage could occur. Turn off the laser before inspecting any delivery system or laser components. Selecta PBL200051, Rev J...
CAUTION - U.S. fedederal law restricts this device to sale by or on the order of a physician. CAUTION - Lumenis medical lasers and laser delivery systems are intended solely for physicians trained in the use of these instruments. Selecta...
Regulatory Compliance 15 Regulatory Compliance Lumenis lasers and delivery systems comply with 21 CFR Chapter 1, Subchapter J, as administered by the Center for Devices and Radiological Health of the US Food and Drug Administration (FDA). CE-labeled devices comply with all appropriate performance standards as specified in Annex II of the Medical Device Directive MDD 93/42/EEC.
The laser has a protective housing that prevents unintended human access to laser radiation above Class I limits. The housing must only be opened by a Lumenis-certified technician. Safety interlocks The protective housing is not designed to be removed by the user during operation or maintenance.
Page 19
Location of Regulatory and Other System Labels 19 3959 West 1820 South Salt Lake City, UT 84104 U.S.A. Lumenis (Germany) GmbH Heinrich-Hertz-Strasse 3 D-63303 Dreieich, Germany +49.6103.8335.0 CAUTION: U.S. LAW RESTRICTS THIS DEVICE TO SALE BY OR ON ORDER OF A PHYSICIAN...
Page 22
22 Safety and Regulatory 3959 West 1820 South Salt Lake City, UT 84104 U.S.A. 0044 LUMENIS (Germany) GmbH (for products shipped under Heinrich-Hertz-Strasse 3 the Medical Device Directive) D-63303 Dreieich, Germany Model name, serial number, and +49.6103.8335.0 manufacturing date...
Page 23
90% @ +55ºC (Danger. Possible explosion hazard if used 14ºF – 131ºF NON-CONDENSING in the presence of flammable anesthetics, oxygen, or nitrous oxide.) Flammable anesthetics and gases warning Nonoperating environmental specifications (on shipping box) Regulatory compliance labels Selecta PBL200051, Rev J...
Trio — a Selecta Duet and a LaserLink S with compatible Lumenis 532 nm photocoagulator Selecta has all of the standard controls and functions of a diagnostic slit lamp and is intended for use in eye examination of the anterior segment, from the cornea epithelium to the posterior capsule.
Page 28
The LaserLink S laser delivery adaptor is used for treatments specifically cleared for the compatible laser photocoagulator. Selecta is shipped directly from the factory to your site. Your local Lumenis representative initially unpacks, inspects, sets up, and installs the Selecta to ensure that it is working properly.
Selecta System Components 29 Selecta System Components Selecta is comprised of the following major components: • a diagnostic slit lamp with parallel binocular optics and five- position magnification • an integrated Nd:YAG laser module that delivers 1064 nm and/or 532 nm wavelengths, depending on the laser model •...
Remote control The remote control is the control panel for the Selecta laser. The remote control allows you to select laser treatment settings, such as laser mode, energy, and aiming beam intensity. The remote control can be placed on the table so that the operator can verify the laser treatment parameters without moving back from the slit lamp binoculars.
CAUTION - Do not use device if slit lamp seems unstable or inadequately secured. Inspect the Selecta components Before connecting the Selecta components, inspect the individual components for dirt, debris, or damage. Check the electrical cables to ensure that they are not frayed or split.
Page 32
32 Operation Slit lamp and table Remote control receptacle (beneath slit lamp table) Connecting the remote control to the slit lamp console Selecta PBL200051, Rev J...
Connection Instructions 33 Connect the footswitch (optional) The Selecta provides a choice of laser activation. By default, the laser is activated by the joystick laser pushbutton. However, many physicians prefer footswitch activation, and Lumenis provides a footswitch that may be used with the system.
To resume treatment, close the treatment room door or reinsert the interlock plug, and press the (ready) button on the remote control. External door interlock receptacle (on back of system) External door interlock plug Connect the interlock plug Selecta PBL200051, Rev J...
Insert the main power plug into the main power receptacle, as shown. Plug the other end into an electrical outlet. Power plug receptacle (on table console, near right rear wheel) Plugging in the main power cable Selecta PBL200051, Rev J...
Installing optional accessories CAUTION - Disconnect or stabilize all accessories during transportation. A variety of optional viewing accessories is available for the Selecta, such as camera and video adaptors, and a co-observation tube. Compatible Selecta accessories are listed in the “Specifications” section of this manual. You may also contact your local Lumenis representative for additional product information.
Page 38
Seat the guide pin into the guide pin slot, and tighten the binocular screw to secure the connection. Binocular Beam assembly Magnification changer splitter screw Binocular screw Beam Beam Binocular splitter splitter dovetail dovetail dovetail mount mount receptacle Secure the beam splitter between the binocular assembly and magnification changer Selecta PBL200051, Rev J...
Page 39
Rotate the beam splitter attachment ring clockwise to secure the connection. Front view Centering groove Beam splitter port Side view Attachment ring Attach the compatible viewing device to the beam splitter (co-observation tube shown) Selecta PBL200051, Rev J...
Attach the LaserLink S fiber optic to the photocoagulator, as described in the LaserLink S operator manual. Disconnect the external door interlock plug from the Selecta and connect the plug to the photocoagulator when the LaserLink S is connected and in use.
Page 41
Connection Instructions 41 THIS PAGE IS INTENTIONALLY BLANK Selecta PBL200051, Rev J...
Laser settings are selected from the remote control. The laser activation mode is user-selectable: either a footswitch or joystick pushbutton may be used. Selecta PBL200051, Rev J...
Page 43
System Basics 43 Slit Lamp Joystick laser pushbutton Remote control Table top Table height control Keyswitch Emergency laser Main power receptacle Power on stop button (near right rear wheel) indicator light Laser footswitch Locking wheels Selecta PBL200051, Rev J...
The standby icon on the remote control illuminates, and the laser is in standby mode. The system will default to minimum settings. Turning off the system Under normal operating conditions, turn the keyswitch to the (off ) position. Selecta PBL200051, Rev J...
46 Operation Emergency off In an emergency, press the red emergency off button to immediately de-energize the system. Emergency off Selecta PBL200051, Rev J...
After disconnecting the system, move the slit lamp table and accessories to the desired site. Step on the wheel brake to lock the wheel, as shown. To unlock the wheel, place your foot under the brake and lift the lever up. Locking the wheels Selecta PBL200051, Rev J...
As shown in the following illustration and described in the following section, the Selecta slit lamp has all of the standard diagnostic slit lamp controls such as illumination intensity, illumination aperture, slit width adjustment and rotation, as well as the typical diagnostic slit lamp adjustments, such as diopter eyepiece adjustments, magnification, binocular focus, and interpupillary distance adjustment.
Page 49
Illumination tower angle scale Microscope arm Joystick locking screw Arm position Illumination angle scale intensity control Instrument base Illumination aperture Aiming beam Instrument base and color filter intensity locking screw selector wheels control Slit lamp controls Selecta PBL200051, Rev J...
Slit lamp controls Eyepieces, diopter adjustment, and interpupillary distance The Selecta slit lamp microscope eyepieces have a 12.5x magnification, with diopter adjustments on both eyepieces of ±5. One eyepiece has crosshairs to aid focusing. The eyepiece positions can be manually adjusted to suit the operator’s interpupillary distance.
100 µm is the nominal setting, as indicated on the control knob. The posterior YAG offset is continuously adjustable from 0 to 350 µm posterior, with detents at 0, 100, 250, and 350 µm. Selecta PBL200051, Rev J...
Error message display activation indicator Energy display Pulse count Burst display Total energy delivered in mJ Inactive in Burst selector Photodisruptor mode Energy control Pulse count & total energy reset Remote control (Photodisruptor-1064 nm screen shown) Selecta PBL200051, Rev J...
Joystick button or footswitch laser activation indicator Energy display Pulse count Total energy delivered in mJ Inactive in SLT Inactive in mode SLT mode Energy control Pulse count & total energy reset Remote control (SLT-532nm screen shown) Selecta PBL200051, Rev J...
PC (external photocoagulator) Mode—remote control screen Options toggle Contrast knob (on top side) (on top side) Laser mode selector PC Mode indicator Illumination mirror indicator Laser mode display Error message display Remote control (PC mode shown) Selecta PBL200051, Rev J...
The Selecta provides a choice of laser activation. By default, the laser is activated by the laser pushbutton on the slit lamp joystick. However, many physicians prefer footswitch activation, and Lumenis provides a Selecta footswitch that may be used with the system.
58 Operation Laser mode If you’ve purchased the Selecta Duo, the laser mode pushbutton lets you select either the YAG or SLT laser mode. The remote control screen will change to reflect all of the treatment settings available in the selected laser mode.
Remote Control System Options Select the options screen The Selecta remote control options screen is where you can set system preferences such as the volume of system beeps, the interface language, and whether the treatment laser is activated by the joystick pushbutton or a footswitch.
Remote Control System Options 61 Language The Selecta system interface can be displayed in either English or icon mode. The language interface button is located on the options screen. In English mode, any informational or error messages are displayed in English and with an error code, such as “E109”.
To decrease the brightness of the LCD screen, press the lower portion of the button. To increase the brightness, press the upper portion of the button. LCD brightness display Increase brightness Decrease brightness Adjusting the remote control screen brightness (Options screen shown) Selecta PBL200051, Rev J...
Remote Control System Options 63 Laser activation mode—Joystick or Footswitch The Selecta provides a choice of laser activation. By default, the laser is activated by the laser pushbutton on the slit lamp joystick. However, the laser can be activated either by a pushbutton on the slit lamp joystick or by a footswitch.
The volume control is located on the remote control options screen. To lower the beep volume, press the lower portion of the button. To increase the volume, press the upper portion of the button. Volume display Increase volume Decrease volume Adjusting the system volume (Options screen) Selecta PBL200051, Rev J...
Refer to the “Specifications” section of this manual for available energy settings. The energy display shows the currently selected energy. Selected energy per burst Energy control Selecting the energy per burst (Photodisruptor mode shown) Selecta PBL200051, Rev J...
Use the reset button to reset the total pulse count and energy displays to zero. Total pulse count Total energy delivered Reset button Total energy and total pulse count displays and reset button (Photodisruptor mode shown) Selecta PBL200051, Rev J...
Set the burst mode using the burst mode selector. The currently selected burst mode is shown in the burst mode display. Burst mode display Burst selector Selecting the photodisruptor burst mode (Photodisruptor screen) Selecta PBL200051, Rev J...
Note the index number of the slit lamp eyepieces relative to the index marks as a future starting point for this procedure. View the target through the eyepieces, and adjust the focus to obtain the smallest spot on the target. Remove the paper target from the headrest. Selecta PBL200051, Rev J...
Set the slit spot to 5 mm. Set the microscope magnification to the maximum setting. Visually verify that the slit spot is centered in the field of view. If the spot is not centered, contact your Lumenis service representative. 5 mm illuminated spot Field of view...
Swing the illumination tower to one side, if necessary, so that it does not obstruct the aiming beam. Position the slit lamp so that the aiming beams converge to form one spot on the paper target. Selecta PBL200051, Rev J...
WARNING - If the aiming beams cannot be made coincident, do not use the laser on patients. Doing so may cause inadvertent damage to nontarget tissue and injury to the patient. Contact your local Lumenis representative. Lock the slit lamp in this position.
Page 72
WARNING - If the laser burn mark and the aiming beams are not coincident, do not use the laser on patients as it may cause serious injury. Contact your local Lumenis representative. Repeat the test as necessary. Selecta...
Operation Instructions 73 Operation Instructions Preoperative Instructions Ensure that the Selecta system is properly connected, as described in the “Connection instructions” section of this manual. Post the “Laser in Use” warning sign outside the treatment room door. Verify that the pre-operative system set up, as described in this manual, was properly performed.
Clean the Selecta, as instructed in the “Maintenance” section of this manual. Cover the Selecta with its dust cover to protect it and ensure that the optical components remain free of dust and other contaminants.
Verify that the delivery system is properly connected. External door interlock If the external door interlock is used in conjunction with a remote switch, verify that the external door interlock plug is inserted in the external door interlock receptacle. Close the interlocked door. Selecta PBL200051, Rev J...
Page 78
Slit lamp binocular eyepieces are not Refer to the Focusing the Slit Lamp sec- properly adjusted. tion of this manual for additional infor- mation for properly adjusting the Slit lamp binocular eyepieces. Selecta PBL200051, Rev J...
Page 79
The LaserLink S is attached and the Remove the LaserLink S. Cannot select the Ready mode Selecta system is in YAG or SLT laser mode. The slit lamp table is in motion. Adjust the table to the desired height and position, and then select the Ready mode.
Page 80
If “I109” displays on the the condition persists, contact your local remote control Lumenis representative. “Energy Low < 20%” or The detected energy level was at least Place the laser in ready mode. If the 20% lower than the level selected.
Page 81
Place the laser in ready mode. If the condi- “Energy Low < 50%” or “C102” displays on the 50% lower than the level selected. tion continues, restart the laser. If the con- dition persists, contact your local Lumenis remote control representative. The laser temperature is too high.
Lumenis-certified service engineer to ensure proper laser performance. Laser repair All laser repairs should be performed by a Lumenis-certified service engineer. For training and information, contact your local Lumenis service representative. Inspecting the laser system components Before and after each use, inspect the laser system components for evidence of dirt, debris, or damage.
You may damage the console and laser system electronics. Cleaning and disinfecting areas of patient contact Selecta laser treatment involves only brief skin contact with the patient, but attention should be given to the possibility of cross-contamination between patients. The areas of skin contact include the slit lamp table, chin rest, headrest and handles.
84 Maintenance Clean the gonioscopy contact lens The Lumenis Selecta contact lenses have a special low-reflectivity coating bonded to the lens, and so must, therefore, be handled carefully. As soon as a lens is removed from a patient’s eye, thoroughly rinse it in cold or warm water to remove salts, mucous and gonioprism solution.
Total length of cable shall not exceed five meters (16 feet). Pin assignments are as follows: Signal Name Signal Description Remote Interlock Connect to switch, normally open Return Connect to switch common None No connection Remote door interlock pin assignments (mating face shown) Selecta PBL200051, Rev J...
Unlock the electrical input module cover by inserting a small flathead screwdriver into the slot. Gently push against the locking tab until the lock releases. Remove the fuse cover. Installing new fuses Electrical module Changing the fuses Selecta PBL200051, Rev J...
Page 87
Replace the two 5 millimeter by 20 millimeter fuses with the appropriate replacement fuses as indicated below: Voltage Configuration Fuse Rating 100-240VAC 3.15A Fuse table Place the cover back onto the module. Gently push against the cover until the locking tab latches. Selecta PBL200051, Rev J...
Be careful not to directly touch the bulb. On the right side of the illumination tower, loosen the bulb housing cover by turning the thumbscrew counter-clockwise. Housing Cover Illumination bulb housing Selecta PBL200051, Rev J...
Page 89
Holding a new bulb by its base, and being careful not to touch the glass with your fingers, insert the prongs securely into the bulb receptacle. Close the cover door and tighten the thumbscrew. Place a new illumination bulb in the receptacle Selecta PBL200051, Rev J...
Calibration questions should be referred to your local Lumenis representative. Disclaimer warning Calibration is a service procedure to be done only by Lumenis-certified service engineers or customers who have taken and passed a Lumenis Service Certification Training course. Adjustment by anyone other than a trained Lumenis service engineer or a certified customer voids any existing manufacturer’s warranty on the instrument.
Selecta Laser System, as this may result in increasedelectromagnetic emissions or decreased immunity to such emissions. WARNING - If the Selecta Laser System is used adjacent to or stacked with other equipment, observe and verify normal operation of the laser system in the configuration in which it will be used prior to using it in a surgical procedure.
92 Maintenance Specifications The Selecta Ophthalmic laser system delivers both a 1064 nm and 532 nm treatment beam; therefore, the laser specifications outlined below for each wavelength are applicable to the Selecta Duet model. Selecta Duet 532 nm laser...
Relative humidity 0 to 90% at 55°C, non-condensing Standard commercial shipping altitude Altitude Vibration Capable of surviving transport by normal commercial recognized air, sea, and land carriers Shock Lumenis shock requirements, based on size and weight Selecta PBL200051, Rev J...
Page 94
0.2, 5, 8.9, 13.3 mm and tear drop (variable 1.7 to 12.2 mm) Green, cobalt blue, Neutral Density (ND) 28% attenuation, and none Illumination filters Fixation lamp Green or red, 12 V Adjustable, vertical travel 50 mm (2 in), with fixation lamp Chin rest Selecta PBL200051, Rev J...
Decontamination of Returned Equipment To comply with United States postal and transportation law, equipment shipped to Lumenis US offices for repair or return must be properly decontaminated with a chemical germicide that is commercially available and cleared for use as a "Hospital Disinfectant." To ensure that all equipment has...
98 Professional Use Instructions Selecta PBL200051, Rev J...
Page 99
Refer to the laser operator manual, Indications for Use section. • Selecta Trio: same indications for use as a Selecta Duet and LaserLink S slit lamp delivery device with a currently cleared Lumenis 532nm photocoagulator...
SLT treatment laser provides a low energy, short pulse of laser light that produces a thermal effect in pigmented cells in the trabecular meshwork. If an optional LaserLink S delivery device is attached to the Selecta system and a compatible Lumenis 532 nm photocoagulator laser, such as the Novus Spectra, the Selecta works strictly as a standard diagnostic slit lamp—...
Page 101
The physician is advised to consult medical journals for information on laser treatment in a particular specialty. The use of a laser is at the physician’s discretion except in cases where the application is contraindicated. Selecta PBL200051, Rev J...
The mechanism of action of the Lumenis Selecta 532nm frequency-doubled, Q-Switched Nd:YAG ophthalmic laser is the selective targeting of pigmented trabecular meshwork cells. The Selecta achieves its intended effect through the use of single laser pulses of short duration and low fluence (energy/area). The...
SLT Precautions The following precautions are suggested when using the Selecta for SLT: • To reduce the risk of damage to non-targeted tissues, the treatment energy setting should be the minimum energy necessary to perform the treatment.
Selecta 1064 nm wavelength acts principally on water in tissue and produces a cutting or disrupting tissue effect. Therefore, the Selecta 1064 nm laser will damage any tissue or structure on which the beam is focused.
Posterior Capsulotomy—1064 nm Nd:YAG 105 Posterior Capsulotomy—1064 nm Nd:YAG The Selecta 1064 nm laser is indicated for disruption of the posterior capsule of the eye (posterior capsulotomy). Contraindications The following represent contraindications for posterior capsulotomy or pupillary membranectomy: • pre-existing ocular pathologies including: –...
Page 106
Patients who exhibit a pressure rise generally return to pretreatment levels within 24 hours but should be carefully followed throughout this period. Persistent IOP elevation occurs in some patients. Clinical estimates are that 2-3% of patients may be treated for secondary glaucoma. Selecta PBL200051, Rev J...
Page 107
Physicians who experience problems with continued pitting should consider ending the treatment. The Selecta aiming beam is offset longitudinally up to 0.1 mm in front of the Nd:YAG beam. Physicians who experience continued focusing problems, which can lead to IOL pitting, should first check the ocular setting of each eye on the slit lamp.
Page 108
• patients with IOLs, particularly posterior chamber lenses close to the posterior capsule, are at risk of IOL pitting or cracking • patients with pre-existing ocular conditions have a greater chance of experiencing postoperative complications Selecta PBL200051, Rev J...
Targeting Focus the unit in accordance with the description provided in the Operation section of this manual. In dual aiming beam systems such as Selecta, the Nd:YAG beam extends beyond the diode beams, so there is an increased probability of Nd:YAG beam impingement of the iris. The diode aiming beam focus spot is longitudinally offset up to a maximum of 350 microns in front of the Nd:YAG beam.
Iridotomy—1064 nm Nd:YAG 111 Iridotomy—1064 nm Nd:YAG The Selecta 1064 nm wavelength is indicated for performing an iridotomy (hole in the iris). Contraindications The following represent contraindications for iridotomy. Pre-existing ocular pathologies including: • eyes with opacities of the media such that the iris cannot be adequately visualized •...
The risk of lens damage during Nd:YAG laser iridotomy will increase if: • laser focusing is inaccurate • laser energy is applied through an already patent iridotomy • laser energy is applied through the pupil directly to the lens Selecta PBL200051, Rev J...
All eyes should be observed with a biomicroscope for bleeding. If bleeding occurs, additional Nd:YAG laser treatment may aggravate it. Further, if bleeding does not stop spontaneously or after applying digital pressure, argon laser photocoagulation of the bleeding site may be necessary. Selecta PBL200051, Rev J...
Nd:YAG laser iridotomy: • patients with chronic uveitis have an increased tendency towards both early and late iridotomy closure • patients with vascularization of the iris or engorgement of iris vessels are at increased risk of bleeding Selecta PBL200051, Rev J...
350 microns in front of the Nd:YAG beam. This offset is adjusted on the laser using the posterior offset control. With the offset control set at 100 microns, Selecta delivers a preset offset equal to 100 microns in air. Energy Minimal effective energy and minimal pulses per burst should be used to reduce risk of damage to nontarget areas.
Page 116
116 Professional Use Instructions Selecta PBL200051, Rev J...
Guidance and Manufacturer's Declaration Electromagnetic Emissions Selecta ophthalmic laser family is intended for use in the electromagnetic environment specified below. The customer or the user of Selecta ophthalmic laser family should ensure that it is used in such an environment.
Page 120
Guidance and Manufacturer's Declaration: Electromagnetic Immunity Selecta ophthalmic laser family is intended for use in the electromagnetic environment specified below. The customer or the user of Selecta ophthalmic laser family should ensure that it is used in such an environment.
Page 121
RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the Selecta ophthalmic laser family system is used exceeds the applicable RF compliance level above, the Selecta ophthalmic laser family system should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating the Selecta ophthalmic laser family unit.
Page 122
Equipment and the Selecta ophthalmic laser family System The Selecta ophthalmic laser family system is intended for use in an electromagnetic environment in which radiated RF distur- bances are controlled. The user of the Selecta ophthalmic laser family system can help prevent electromagnetic interference by...
The undersigned also agrees to reimburse Lumenis for any costs incurred in cleaning the enclosed equipment, in the event said item(s) is/are received by Lumenis in a contaminated condition.
Need help?
Do you have a question about the SELECTA and is the answer not in the manual?
Questions and answers
WHAT IS SAVER MODE ON MY LASER
Saver Mode, referred to as CaseSaver™ Mode, is a feature on the LUMENIS SELECTA laser that allows continued operation if one of the laser wavelengths becomes inoperable. If the unit has more than one wavelength, the remaining operational wavelengths continue to function, enabling treatment to proceed until the unit is serviced.
This answer is automatically generated