Download Print this page

YOKOGAWA FLXA21-P1 Technical Notes page 4

Programming custom buffer tables

Advertisement

10.)
Depending on which buffers you wil
be using in your application, click on the
diamond next to the User defined buffer
that needs to be changed.
a. User defined Buffer 1, is the buffer
table for pH 4.01;
b. user defined Buffer 2, is the the
buffer table for pH 6.87;
c. and user defined Buffer3, is the
buffer table for pH 9.18.
For this example we will choose to change the values within the User
Deinfed Buffer 2, pH 6.87, because this is the most commonly choosen
table to change values to reflect pH 7.0.
11.)
Select the diamond next to Enter
Values
Tech Note: TNA1403
Date: February 5, 2014