Air filter
Standard:
Hypoallergenic:
Aircraft use
ResMed confirms that device meets the Federal Aviation Administration (FAA) requirements (RTCA/DO-160, section 21,
category M) for all phases of air travel.
Operating pressure range
APAP, CPAP:
Supplemental oxygen
Maximum flow:
Pneumatic flow path
Design life
Device, power supply unit:
Cleanable water tub:
Standard water tub, air tubing:
General
The patient is an intended operator.
Humidifier performance
The following settings have been tested at 71.6°F (22°C) ambient temperature:
Mask Pressure
RH output %
cm H
O
Setting 4
2
4
85
10
85
20
85
1
AH - Absolute Humidity in mg/L
2
BTPS - Body Temperature Pressure Saturated
Air tubing
Air tubing
Material
SlimLine
Flexible plastic
Standard
Flexible plastic
Notes:
 The manufacturer reserves the right to change these specifications without notice.
 Do not use electrically conductive or antistatic air tubing.
 The temperature and relative humidity settings displayed are not measured values.
Material: Polyester non woven fiber
Average arrestance: >75% for ~7 micron dust
Material: Acrylic and polypropylene fibers in a polypropylene
carrier
Efficiency: >98% for ~7-8 micron dust; >80% for ~0.5 micron
dust
4 to 20 cm H
O
2
4 L/min
1. Flow sensor
2. Blower
3. Pressure sensor
4. Mask
5. Air tubing
6. Water tub
7. Device
8. Inlet filter
5 years
2.5 years
6 months
Nominal system output AH
Setting 8
Setting 4
100
6
100
6
90
6
Length
6' (1.8 m)
6'6" (2 m)
1
2
, BTPS
Setting 8
>10
>10
>10
Inner diameter
0.6" (15 mm)
0.75" (19 mm)
English