Ohmeda Ohio 2000 Series Operation And Maintenance Manual page 31

Infant warmer systems
Table of Contents

Advertisement

11/09/93
+
Heat Reflecting Patch
with Foil Side Facing Up
White Paper
Protector
Patient Probe
(Place the metal side in contact
with the infant's skin)
Note: Avoid placing
excessive strain
on the skin temperature
probe lead. Always
remove the probe by
grasping the plug at the
panel. Do not pull on the
probe lead.
5.
Servo Mode Operation. (all models except the 2001 International)
Note: The skin temperature probe must be properly attached before
starting servo mode operation. Refer to step 4.
You must select a servo control temperature setting when the
Warmer is used in the servo mode for the first time. You are
prompted to make this setting with an operator prompt tone and the
flashing of the control temperature display.
3-10
6600-0194-000
n
!
Figure 3-2
Skin Temperature
Probe attachment
Figure 3-3
Temperature Probe
connection to the
Display Module
CAUTION: Always remove the probe from the patient by grasping
and removing the heat reflective patch first, then remove the probe
from the patient or the patch. Always remove the probe from the
Warmer by grasping the plug at the panel. Placing excessive strain
on the skin temperature probe lead can damage the probe.
3/0peration
Connect the skin temperature probe to the Warmer by plugging the
probe connector into the left side of the display module as viewed
from the front. (See Figure 3-3.)