Ohmeda Ohio 2000 Series Operation And Maintenance Manual page 27

Infant warmer systems
Table of Contents

Advertisement

11109/93
*
15
minutes
1
minute
Cannot be
5
minutes
12 minutes
15 minutes
Cannot be silenced
Off
Off
Off
*
Off
Off
3-6
6600-0194-000
.5 and 2°C by qualified service personnel.
Alarm
Silence Period
1
minute
1 minute
Heater
Off
Off
5
minutes
§
Can be set to between
* Not applicable to model 2001 International
alann.
t
Preceded by a 30 second operator prompt tone.
* Error code in elapsed time display indicates the cause of the
*
Heater output is dependent on the Patient Temperature and Control Temperature
settings in the servo mode, and the % Power setting in the manual mode.
l
t
Alternating Two-Tone
silenced
Alternating Single Tone t
Alternating Single Tone t
after 12 minutes
Alternating Two-Tone
after 15 minutes
Power Failure
Alternating Single Tone
*
System Failure* *
Heat Off
Check Patient
Alternating Single Tone t
Alternating Two-Tone
30°C
6of
Control temperature*
Patient Temperature
is less than
Z°C
2'C from
Control temperature*
Alternating Single-Tone t
Patient Temperature
between 1 and
42'CS
Alternating Two-Tone t
Patient Temperature
more than
Failure*
Alternating Two-Tone t
Patient Temperature
is greater than
- Warmer Alarms
Alarm
Condition
Alarm Sound
Probe
3/0peration
Check Patient Alarm:
1.
Manual Mode: The check patient alarm activates in the manual mode
of operation when the heater has been energized at a power level
greater than 25% for 12 minutes. The alternating single tone alarm is
activated for this condition. Pressing the alarm silence switch silences
this alarm and resets the timer for another 12 minutes of operation. If
this alarm is not silenced within 3 minutes of the check patient alarm
occurring, an alternating two tone alarm is sounded and the heater is
deactivated until the alarm silence switch is pressed.
2.
Servo Mode: The check patient alarm activates in the servo mode of
operation if the heater has been at the 100% power level for 12 con-
tinuous minutes. The alternating single tone alarm is activated for this
alarm. Pressing the alarm silence switch silences this alarm and resets
the timer for another 12 minutes of operation. If this alarm is not
silenced within 3 minutes of the check patient alarm occurring, the
alternating two tone alarm is sounded and the heater is deactivated
until the alarm silence switch is pressed.
Power Failure Alarm:
A battery operated power failure alarm activates when the external power
source fails or is accidentally disconnected. The battery also provides
power to an electronic memory to recall previous control temperature
settings for approximately 10 minutes, when the battery is fully charged.
Table 1