11/09/93
tiltable bed provides a smooth, dampened motion to
avoid disturbing the patient. * A chest drainage hanger attached to the
bed and rear cross bar extrusion provides a convenient location to hang
chest drainage devices. *
1.6 Bassinet (3500 Series)
The bassinet unit can be detached and rolled away from the model 3500
warmer unit by pulling up on the locking lever knob located at the top
rear of the bassinet (see Figure l-l). The bassinet includes:
1.
Transparent locking side panels which fold down for easy patient
access and are removable for cleaning.
l
Not available on the model 3500 wanner
l-2
6600-0194-000
tiltable
bed platform allows Trendelenburg and Fowler positioning. The hydrau-
lic system for the
l/Description
1.4
The heater assembly consists of a radiant heater, parabolic reflector,
observation light, and a visual alarm light. The parabolic reflector focuses
radiant energy on the bed surface, minimizing energy loss due to scatter-
ing and providing an even field of radiant heat over the bed surface. The
observation light provides intense light for procedures. The alarm light is
located on the end of the heater assembly for ease of viewing. The entire
heater assembly rotates to the side for X-ray procedures and for observa-
tion lamp replacement.
Control Unit
The control unit contains the electronic circuits and controls used to
operate the radiant heater and the observation light. The control unit
performs regular self checks during its operation including failure diag-
nostics.
Either manual or servo mode of operation may be selected. In the manual
mode of operation, select the level of radiant heat output as indicated by
the % power display on the control panel. The control circuit then main-
tains the selected level of radiant heat. The manual mode has a preheat
setting which allows the Warmer to be preheated. In the servo mode of
operation, select the patient's control temperature. A skin temperature
probe is used to monitor the patient skin temperature. The control
system modulates the radiant heat to maintain the patient at the se-
lected control temperature. The patient's skin temperature is continu-
ously displayed.
Alarms activate to alert the operator of a low or high patient tempera-
ture, a skin temperature probe failure, a power failure, equipment failure
or a check patient prompt.
1.5 Bed Platform
Warmers with beds include a mattress and transparent side panels. The
side panels fold down for easy access to the patient and can be removed
for cleaning. The X-ray cassette tray (located in an opening beneath the
bed when installed) pulls out for insertion of X-ray cassettes and allows
X-rays to be taken without moving the patient. * Markings located along
the side panels allow easy location of the cassette in the X-ray tray
relative to the position of the patient on the warmer bed.*