INSTALLATION

Inspect the controller for any shipping damage.
Should the controller be connected to a host computer
via the-RS-232 interface, a suitable cable must be
prepared.
In the following paragraphs are detailed the input/output
signals.
The box installed into the customer system must be
positioned so that cold air (forced or natural convection)
can flow around.

Pump Connector

The signals of J3 connector are the following:
Pin C 80 Vac 3-phase output to pump motor stator
(phase T).
Pin D 80 Vac 3-phase output to. pump motor
stator (phase S).
Pin B 80 Vac 3-phase output to pump motor stator
(phase R).
Pins A/F Pump temperature sensor.
Pin E Ground
NOTE

I/O Specifications

START/STOP:
- START command
- STOP command
Analog output:
- Output impedence
- Minimum load:
Normal operation signal:
- Open Collector
Current rating
Low speed command:

9-pin "D" Type Connector Pin Assignement

Pin
Description
number
1
Start/Stop input: close to pin 5/6 to start the
pump
2
Pump in Normal output: closed to pin 5/6 when
pump speed is higher than 80% of full speed
3
Earth (Ground)
4
Analog output proportional to pump speed
(positive)
5-6
Electrical supply (0 V)
9
Low speed input: close to pin 5/6 to select Low
Speed mode
7-8
Electrical supply (75 V) (positive)
*
Minimum speed reading
in STOP condition = 100 Hz (6 KRPM)
4
TECHNICAL INFORMATION
Low <0.8 Vdc
High 4 to 15 Vdc
0 to 10 Vdc (proportional to
*
speed)
(0 to 10 V ≡ 0 to 100% speed)
0.1 Ω
2 K
(5 mA)
Speed <80%: OFF (pull-up
to 15 Vdc)
Speed >80%: ON (<0.8 Vdc)
60 mA max
Low (<0.8 Vdc)
87-900-944-01 (C)