Maintenance

Multiple hazards. Only qualified personnel must conduct the tasks described in this section of the
document.
Maintenance schedule
The maintenance schedule shows minimum intervals for regular maintenance tasks. Perform
maintenance tasks more frequently for applications that cause electrode fouling.
Note: Do not disassemble the probe for maintenance or cleaning.
Maintenance task
Clean the sensor
Inspect the sensor for damage
Replace the sensor cap
Clean the sensor
Clean the exterior of the sensor with a soft, wet cloth.
Note: If the sensor cap must be removed for cleaning, do not expose the interior of the cap to direct sunlight for
extended periods of time.
Change the sensor cap
Potential explosion hazard. The sensor setup cap is not rated for hazardous location use.
Replacement sensor caps and setup caps are shipped with installation instructions. Refer to the
included instructions to change the cap. After the new sensor cap is installed, make sure that the lot
number on the sensor cap is the same lot number that is read by Modbus. The sensor uses the
calibration information from the sensor cap.
For best performance and accuracy, replace the sensor cap:
• Every two years, or more frequently if necessary
• When routine inspection shows significant erosion of the sensor cap
mS/cm
PSU
19
13
20
13
21
14
22
15
23
15
24
17
25
17
26
18
W A R N I N G
W A R N I N G
mS/cm
PSU
30
21
31
22
32
22
33
23
34
24
35
25
36
25
37
26
Recommended minimum frequency
90 days
90 days
2 years or more frequently, if necessary
mS/cm
PSU
42
30
44
32
46
33
48
35
50
37
52
38
54
40
English 9