Storage temperature

Degree of protection

Shock resistance

Vibration resistance

Interior cleaning

Electromagnetic
compatibility (EMC)

Medium temperature range

Medium pressure range
(nominal pressure)

Density

Pressure-temperature
ratings
22
US units
T
a
[°F]
T5
[212 °F]
140
194
The minimum temperature of the medium is –40 °C (–40 °F).
The minimum ambient temperature is –40 °C (–40 °F).
–40 to +80 °C (–40 to +176 °F), preferably at +20 °C (+68 °F)
As standard: IP67, type 4X enclosure
As per IEC/EN 60068-2-31
Acceleration up to 1 g, 10 to 150 Hz, based on IEC/EN 60068-2-6
• Cleaning in place (CIP)
• Sterilization in place (SIP)
Observe the maximum medium temperatures →  22
According to IEC/EN 61326
For details, refer to the Declaration of Conformity.

Process

Sensor
–40 to +125 °C (–40 to +257 °F)
Cleaning
+150 °C (+302 °F) / 60 min for CIP and SIP processes
Seals
No internal seals
max. 40 bar (580 psi), depending on process connection
0 to 5 000 kg/m
3
(0 to 312 lb/cf)
The following pressure-temperature ratings refer to the entire device and not just the process
connection.
Maximum medium temperature T
T4
T3
[275 °F]
[392 °F]
257
257
Dosimass
m
T2
T1
[572 °F]
[842 °F]
257
257