Gassing Accessories - Honeywell Searchpoint Optima Plus Operating Instructions Manual

With optional hart output
Hide thumbs Also See for Searchpoint Optima Plus:
Table of Contents

Advertisement

MAN0905_Issue 7_07/17
Storm Baffle (2108B0280)
Provides increased protection
against torrential rain and driven
sea-spray. Recommended for
exposed installations in maritime
locations or areas where monsoon
or tropical rain conditions are
experienced. Also provides protection
against steam and can be used in
unfiltered ducts.
This accessory will increase the
detector's response time.

2.6 Gassing Accessories

Searchpoint Optima Plus has a range of accessories designed to assist with gas testing.
Calibration Cap (2108B0272)
Inlet
Outlet
Used to achieve accurate calibration
of Searchpoint Optima Plus. Requires
all weather protection to be removed
except dust barrier.
Caution: The Calibration Cap and the Gassing Cover must be removed after use.
2. Introduction
Gassing Cover (2108D0258)
15
Searchpoint Optima Plus with optional HART
Dust Barrier (2108B0259)
Prevents the ingress of dust or oil
mist and protects the optics in very
wet or exposed locations. Fits under
the Standard Weather Protection.
This accessory will increase the
detector's response time.
Used to apply gas to Searchpoint
Optima Plus in the field. Suitable for
functional checking. Fits over the
standard weather protection. (For
calibration, use the Calibration Cap).
output
®
Operating Instructions
2108M0550

Hide quick links:

Advertisement

Table of Contents
loading