Calibration control functions
You first need to select the H
Calibration -> Please enter code -> Calibrate probe -> Choose component -> H
The following screen appears:
Cal after install
Start zero cal.
Start MR cal.
Input TC parameters
Now you can adjust the sensor as described in the following sections.
The following sequence must be observed when calibrating the sensor:
1. Enter correction factors for temperature compensation of zero point
2. Enter correction factors for temperature compensation of sensitivity
3. Caibrate zero point of the H
4. Caibrate measuring range of the H
5. Enter installation date.
Steps 1, 2 and 5 are only carried out when installing a new sensor.
Enter correction factors of temperature compensation
You use this function to enter the temperature compensation parameters for calibration of the zero point and sensitivity.
To do this, navigate as follows: Calibration -> Please enter code -> Calibrate probe -> Choose component -> H
Calibrate TC parameters. The following screen appears:
TK Nullpkt. Eingabe
TK Empf.
Eingabe
Temperature compensation of zero point
To do this, proceed as follows:
1. Position the cursor on the first line
2. Then press the ENTER.
button. The following screen appears:
A: -3.0817e+1
B: +2.2547e+0
C: -1.1450e-1
D: +2.8000e-3
You can now view the factors of the TC parameters for the zero point, and change them if necessary.
Supplement 1 for Operating Instructions C79000-G52xx-C216-03
A5E02769382-03, 03/2012
S sensor in order to adjust it. To do this, navigate in the operator menu as follows:
2
S sensor
2
S sensor
2
S.
2
S ->
2
9