Operating mode
In operating mode, you can view device parameters or adjust the device and set the parameters.
This supplement only describes the functions relevant for operating the H
control functions can be found in the standard operating instructions of the ULTRAMAT 23 (C79000-G52xx-C216-03).
Press the ENTER key in measuring mode to start the operating mode. The main menu appears with the follow options.
Analyzer status
Calibration
Parameters
Configuration
Diagnostic control functions
H
S measuring capacity
2
To specify the value of the H
Diagnostics -> Analyzer status -> Measuring reserve - > H2S Measuring reserve.
The following screen appears:
H2S Sensor status
Reading
: 747 ,OO nA
Maint.Req.: 373,50 nA
Fault
:
74,70 nA
The H
S sensor ages with increasing operating time, thereby steadily decreasing its sensitivity. It has almost reached the
2
end of its service life (maintenance requirement) when it has lots 50% of its sensitivity. The sensor must be replaced. If the
sensitivity drops below 10%, error message S30 (sensitivity of H
Diagnostic values
The displayed diagnostic values provide important information for troubleshooting and calibration tasks. To view the
diagnostic values of the H
Diagnostics -> Diagnostic values -> Probe diagnostic values -> Choose component -> H
The following screen appears:
Sensorsig.: 4886
H2S
:
6.5 vpm
Inst. Date: 05.11.2011
Inst Curr : 747
This dialog shows the diagnostic values of the H
● Probe current is the current pressure-corrected current of the H
● H
S is the current hydrogen sulphide measured value in vpm. Negative values are also possible here.
2
● Date is the installation date of the H
● Installation current is the pressure-corrected current per vpm H
8
S measuring capacity, navigate as follow in the operator menu:
2
S sensor, navigate as follow in the operator menu:
2
nA
nA
S sensor. They have the following meaning:
2
S sensor
2
S sensor. A complete overview of all the device's
2
S sensor outside tolerance limits) is output.
2
2
S sensor in nA
2
S of the sensor at the time of installation.
2
Supplement 1 for Operating Instructions C79000-G52xx-C216-03
S.
A5E02769382-03, 03/2012