SETTINGS AND CALIBRATION
Click Measure and Aqua4Plus 2.0 will take 10 readings and
display the average. Watch for stability while Aqua4Plus 2.0
is measuring to ensure an accurate calibration.
Click Accept and Aqua4Plus 2.0 will calculate a new slope
and offset, click Apply to accept the new slope and offset
and Aqua4Plus 2.0 will provide a real time reading to verify
calibration was successful.
Groundwater Elevation
Position the sensor in its desired location and ensure all
data has been uploaded and erased from the sensor before
proceeding with calibration.
Once positioned connect to sensor and select the calibration
button, followed by selecting the Groundwater Elevation
option. Double check measurement units selection before
proceeding.
Enter your current Groundwater Elevation reading in the
Ref. Point box, making sure to match measurement units.
Click Measure and Aqua4Plus 2.0 will take 10 readings and
display the average. Watch for stability while Aqua4Plus 2.0
is measuring to ensure an accurate calibration.
Click Accept and Aqua4Plus 2.0 will calculate a new offset,
click Apply to accept the new offset and Aqua4Plus 2.0
will provide a real time reading to verify calibration was
successful.

Staff Gauge

Position the sensor in its desired location and ensure all
data has been uploaded and erased from the sensor before
proceeding with calibration.
Once positioned connect to sensor and select the
calibration button, followed by selecting the Staff Gauge
option. Double check measurement units selection before
proceeding.
Enter your current Staff Gauge reading in the Ref. Point
box, making sure to match measurement units.
Click Measure and Aqua4Plus 2.0 will take 10 readings and
display the average. Watch for stability while Aqua4Plus 2.0
is measuring to ensure an accurate calibration.
Seametrics • 253.872.0284
Click Accept and Aqua4Plus 2.0 will calculate a new offset,
click Apply to accept the new offset and Aqua4Plus 2.0
will provide a real time reading to verify calibration was
successful.

Removing calibration setup values

To return to factory default simply enter an offset of 0 and
slope of 1 in the Offset and Slope boxes:
Click Apply to confirm settings and Aqua4Plus 2.0 will
provide a real time reading to confirm.
Adjusting for specific gravity
You man enter the specific gravity of your fluid in the Slope
field to adjust for specific gravity when needed. Click Apply
to confirm settings and Aqua4Plus 2.0 will provide a real
time reading to confirm.