2.3.2 SEM Scanner 200 Display

SEM
Value
 (Delta)
Status
indicator
Item
Action Button
Battery Gauge
 (Delta)
SEM
Pressure Indicator
Status Indicator
Table 1. Legend for Error! Reference source not found.
A SEM Value of zero (0.0) shown on the display screen represents when the device is not
measuring or when SEM readings are cleared out from the screen.
Figure 4. SEM Scanner 200 Display (V3.60)
Clears SEM readings and turns the Scanner unit ON/OFF.
Battery icon indicating current battery state of charge
Calculated difference between the minimum and maximum SEM
Values in the set of readings taken
Display of SEM Value for the last completed reading
Bars indicating the applied pressure while taking a SEM reading. The
color changes from yellow to green to yellow as applied pressure
increases. Green bars indicate the correct pressure is applied to take
the SEM reading and the SEM Value will show on the screen.
Number of acquired readings, status indicator, or error messages
Page 8 of 36
OTH-SEM-IFU-
Function
MULTI-LANG-0123 Rev D,
ENGLISH (EU)
Action button
Battery gauge
Pressure
Indicator
Yellow (top bars)
– too much
pressure
Green (middle bars)
– correct pressure
Yellow (bottom bars)
– too soft pressure