BBI SEM Scanner 200 Instructions For Use Manual page 33

Table of Contents

Advertisement

Immunity
IEC 60601
Test
Test Level
Power
30 A/m
frequency
(50/60 Hz)
magnetic field
IEC 61000-4-8
Conducted RF
3 Vrms
IEC 61000-4-6
150 kHz to 80
MHz
Radiated RF
3 V/m
IEC 61000-4-3
80 MHz to 2.7
GHz
Compliance
30 A/m
3 Vrms
3 V/m
Table 11. Electromagnetic Immunity
Page 33 of 36
OTH-SEM-IFU-
Electromagnetic Environment -
Guidance
Power frequency magnetic fields should be
a level characteristic of a typical location in
a typical commercial or hospital
environment.
Portable and mobile RF communications
equipment should be used no closer to any
part of the SEM Scanner 200 system,
including cables, than the recommended
separation distance calculated from the
equation applicable to the frequency of the
transmitter.
Recommended separation distance:
= 2 .
d
1
P
150kHz to 80 MHz
= 2 .
d
1
P
80 MHz to 800 MHz
=
d
2
3 .
P
800MHz to 2.7 GHz
where P is the maximum output power
rating of the transmitter in watts (W)
according to the transmitter manufacturer
and d is recommended separation distance
in meters (m).
Field strengths from fixed RF transmitter, as
determined by an electromagnetic site
survey a, should be less than the
compliance level in each frequency range b.
Interference may occur in the vicinity of
equipment marked with the following
symbol:
MULTI-LANG-0123 Rev D,
ENGLISH (EU)
January 2018