Cleaning And Disinfection; Cleaning And Disinfection Overview; Scanner Cleaning And Disinfection Instructions - BBI SEM Scanner 200 Instructions For Use Manual

Table of Contents

Advertisement

4.

Cleaning and Disinfection

Cleaning and Disinfection Overview

BBI has conducted testing using Metrex CaviWipes
low-alcohol-based wipes are capable of cleaning and disinfecting the SEM Scanner against the
following 5 micro-organisms with kill times of less than 3 minutes:
Mycobacterium terrae
Staphylococcus aureus
Pseudomonas aeruginosa
Escherichia coli
Klebsiella pneumoniae
The SEM Scanner should never be exposed to the agents listed in the Warranty and Disclaimer
section. Use of these agents will void the product warranty.

Scanner Cleaning and Disinfection Instructions

To properly clean and disinfect the SEM Scanner, wipe it for at least 1 minute and 45 seconds
and allow it to dry for at least 2 minutes.
Follow the steps below to clean and disinfect the SEM Scanner:
1. Obtain 3 cleaning/disinfectant wipes. Use the first wipe for 45 seconds to clean the
scanner. Wipe all surfaces of the SEM Scanner, making sure to clean the crevices on
the sides of the device as well (Figure 15-17).
TM
(<20% alcohol), and determined that the
Page 28 of 36
MULTI-LANG-0123 Rev D,
OTH-SEM-IFU-
ENGLISH (EU)
January 2018