BBI SEM Scanner 200 Instructions For Use Manual page 25

Table of Contents

Advertisement

For the heel(s), BBI recommends taking SEM Scanner 200 readings at the four (4) points as shown
in Figure 13 and described below. Readings at these four points were taken for the clinical study
described in Section 3.6. SEM Scanner readings do not need to be taken in numerical sequence.
If, however, one of the Reading points is calloused, do not take a reading at that spot. The SEM
Scanner will produce an SEM Δ reading after only three (3) readings.
Reading #1: Back of the heel around the calcaneus. Avoid
placing the SEM Scanner on the achilles tendon.
Reading #3: Medial side of the heel. Medial side is the same
side as the big toe (first toe). Avoid placement of the SEM
Scanner on the medial malleolus.
Figure 13. SEM Scanner 200 readings over and around the heel
Reading #2: Lateral side of the heel. Lateral side is the same
side as the little toe (fifth toe). Avoid placement of the SEM
Scanner on the lateral malleolus
Reading #4: Heel pad (sole of the heel) from the back edge of
the heel.
Page 25 of 36
MULTI-LANG-0123 Rev D,
OTH-SEM-IFU-
ENGLISH (EU)
January 2018