6.2. Electromagnetic Immunity

Immunity
IEC 60601
Test
Test Level
Electrostatic
± 8 kV contact
discharge (ESD)
± 15 kV air
IEC 61000-4-2
Electrical fast
± 2 kV for
transient/burst
power supply
IEC 61000-4-4
lines
± 1 kV for
input/output
lines
Surge
± 1 kV line(s) to
IEC 61000-4-5
line(s)
± 2 kV lines to
earth
Voltage dips,
<5% UT (>95%
short
dip in UT) for
interruptions,
0.5 cycle
and voltage
40% UT (60%
variations on
dip in UT) for 5
power supply
cycles
input lines
70% UT (30%
IEC 61000-4-11
dip in UT) for 30
cycles
<5% UT (>95%
dip in UT) for 5s
Compliance
Electromagnetic Environment -
Guidance
± 8 kV contact
Floors should be wood, concrete or
ceramic tile. If floors are covered with
± 15 kV air
synthetic material, the relative humidity
should be at least 30%.
± 2 kV for
Mains power quality should be that of a
power supply
typical commercial or hospital environment.
lines
n/a (unit does
not contain any
signal, control,
or telcom lines)
± 1 kV line(s) to
Mains power quality should be that of a
line(s)
typical commercial or hospital environment.
± 2 kV lines to
earth
<5% UT (>95%
Mains power quality should be that of a
dip in UT) for
typical commercial or hospital environment.
0.5 cycle
If the user of the SEM Scanner 200 charging
system requires continued operation during
40% UT (60% dip
power mains interrupts, it is recommended
in UT) for 5
that the SEM Scanner 200 system be
cycles
powered from an uninterruptible power
70% UT (30% dip
supply or a battery.
in UT) for 30
cycles
<5% UT (>95%
dip in UT) for 5s
Page 32 of 36
OTH-SEM-IFU-
ENGLISH (EU)
MULTI-LANG-0123 Rev D,
January 2018