3.3. Use of Device

The SEM Scanner 200 should be properly cleaned and disinfected after it is used on a patient.
The SEM Scanner 200 must be fully dry after cleaning and disinfection prior to use. See Section
4 for detailed cleaning and disinfection instructions.
Follow the steps below to start a session:
1. Remove the SEM Scanner 200 from the charging mat.
2. Verify the scanner has no visible damages or sharp edges and that the electrode is sealed
against the base.
Do not use the device if the electrode seal is broken or any damage or sharp edges are
found.
3. Turn the SEM Scanner 200 on by pressing the Action Button
one second (Figure 8).
3.1 After turning on the SEM Scanner 200,
➢ an audio tone will sound and the initialization screen will be displayed (Figure
9).
➢ It is important that the SEM Scanner 200 electrode remains untouched during
initialization. See display, "Do Not Touch Electrode Until Device Ready".
➢ If the electrode is touched during initialization, turn off the device and turn on
the device to re-initialize.
Figure 8. Starting a Session
Page 15 of 36
OTH-SEM-IFU-
ENGLISH (EU)
for approximately
MULTI-LANG-0123 Rev D,