Loading Slides Into A Magazine; Processing Slides - CellaVision DM1200 User Manual

Hide thumbs Also See for DM1200:
Table of Contents

Advertisement

Note:
The order ID may be up to 24 characters (ASCII) including spaces. No leading
spaces are allowed. The order ID may not begin with "PB", "BFS", or
"ERR".
For more information on order ID barcodes, see A.3.5 Barcodes.
When processing a slide the system searches for order data in the following
order:
1. In the database, unsigned orders with the same order ID.
2. In the database, pending orders.
3. From the LIS.
4. Default values defined in the Analysis tab in Settings.

2.5 Loading slides into a magazine

Load the slide into the magazine with the barcode upwards and towards the
open end. 12 slides can be loaded into a magazine. The slide positions in the
magazine are numbered 1-12 from the bottom up. Make sure that slides are
fully inserted in the magazine before using the magazine.
Note:
A magazine can be used 100 times. The CellaVision DM Software will display a
message when a magazine has been used more than 100 times.

2.6 Processing slides

View the processing in the animated tutorial.
Open the animated tutorial, go to Help/Tutorial.
User's Manual
CellaVision
Caution!
Use only clipped/round/beveled corner slides. Failing to do so may
cause jams and excessive wear on magazines and the system.
Caution!
If you want to process an already processed slide, carefully wipe the oil
off and make sure the barcode is clean and undamaged.
Click Analyze.
PM-10829-01 2015-07-10
®
DM1200
Operating procedures
13

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the DM1200 and is the answer not in the manual?

Table of Contents

Save PDF