CellaVision DM1200 User Manual page 55

Hide thumbs Also See for DM1200:
Table of Contents

Advertisement

PLT result:
1. Click Calculate PLT Result in the PLT Count panel.
2. If you wish to report the PLT results calculated from the number of PLTs per
HPF, you have two choices:
• Select Calculated estimate to report a concentration. The estimate is
• Select Calculated level to report one of four levels: Significantly
3. If you wish to override the calculated PLT results, select Manual level and
choose one of the four levels.
Note:
In the calculations, several decimals are used. The presented results are
truncated to 1 decimal.
Note:
It is not possible to calculate a PLT result for images captured on a
CellaVision
User's Manual
CellaVision
calculated as [Average PLTs/HPF value] x [PLT estimate factor].
decreased, Decreased, Normal or Increased.
Important!
You must determine and enter a PLT estimate factor before you can use
the system to calculate PLT concentration. By default, the PLT estimate
factor is set to 0.
For information on how to set the PLT estimate factor, see To calculate
and set the PLT estimate factor in 9.3.5 PLT settings.
®
Image Capture System.