Figure 7 To Use the Compensation Signal
4.
Set the attenuation on the probe to 10X. Then press AUTO.
5.
Observe the waveform on the screen. In normal conditions, the square
waveform is displayed, as shown in the figure below. If the square
waveform actually shown does not match that in the figure below, please
follow the procedures for "Probe Compensation" in the next section.
Figure 8 Probe Compensation Signal (Square Waveform)
6.
Use the same method to inspect other analog channels.
WARNING
To avoid electric shock, please ensure that the insulated wire of
the probe is in good condition, and do not touch the metallic part
of the probe when the probe is connected to high voltage source.
Tip
The probe compensation signal can only be used for probe compensation
adjustment and cannot be used for calibration.
DS4000E Quick Guide
RIGOL
English
Probe Compensation
Signal Output Terminal
Ground Terminal
15