Pen type, separate electrode SOIL PH METER Model : PH-220S Your purchase of this SOIL PH METER marks a step forward for you into field precision measurement. Although this METER is a complex and delicate instrument, its durable structure will...
1. FEATURES * Complete set with pen pH meter, soil pH electrode, pH 4.0 /pH 7.0 standard buffer solution. * Spear Tip pH electrode ( included ), ideal for penetration pH measurement into soil, meats, cheese and other type of measurements requiring sample penetration.
Page 4
Accuracy ± 0.07 pH(pH5 to pH 9) ± 0.1 pH(pH4 to pH4.9, pH9.1 to pH10) ± 0.2 pH(pH1 to pH3.9, pH10.1 to pH13) * 23± 5℃ * After calibration. Input 10^12 ohms. Impedance Temperature Not necessary to make adjustment. Compensation Sampling Approx.
Page 5
Standard Instruction Manual......1 PC. Accessories Soil pH electrode, PE-06HD..... 1 PC. pH 4.0 buffer solution, PH-04A..1 PC. pH 7.0 buffer solution, PH-07A..1 PC. Optional * Soft carrying case with sash Accessories ( 210 x 110 x 55 mm ), Model : CA-05A * Hard carrying case ( 280 x 195 x 65 mm ), Model : CA-06...
4. pH CALIBRATING PROCEDURE 4-1 Calibrating Consideration The pH meter already calibrated by mV signal that simulated from the ideal pH ELECTRODE mV output ( base on 25 environment ). However due to (a) An ℃ ideal electrode will produce 0 mV at pH 7.00, but most electrodes are slightly off.
Page 9
pH 4 calibration 1)Rinse the electrode in distilled water. Place electrode into buffer pH 4.00 solution. 2)All the calibration procedures are same as the above pH calibration, except : a. Change the buffer solution to pH 4.00 ( PH-04A ). b.
5. MEASURING PROCEDURE After the meter along the pH electrode is calibrated ( section 4-3 ) , the unit is ready for measuring. 1)Connect the pH electrode plug ( 3-10, Fig. 1 ) to the BNC socket ( 3-5, Fig. 1 ) and place electrode into the measured solution.
Display reverse During the measurement, press the " Display reverse button " ( 3-4, Fig. 1 ) once will reverse the display direction for the user convenience. * Press the " Display reverse button " once again will return to the normal display. Auto power off The meter is built the "...
7. OPTIONAL ACCESSORIES SPEAR TIP pH ELECTRODE Model : PE-06HD Spear tip pH electrode is ideal for penetration pH measurements into meats, cheese, soil and other types of measurements requiring sample penetration. pH 7.00 buffer solution Model : PH-07A pH 4.00 buffer solution Model : PH-04A CARRYING CASE Model : CA-06 Hard carrying case.