Sd card real time data logger dissolved oxygen meter (33 pages)
Summary of Contents for Lutron Electronics SL-4011
Page 1
SOUND LEVEL METER Model : SL-4011 Your purchase this SOUND LEVEL METER marks a step forward for into field precision measurement. Although this METER is a complex delicate instrument, durable structure will allow many years of use if proper operating techniques are developed.
1. FEATURES * Large LCD display, easy to read. * Main functions are designed to meet IEC 61672 class 2. * A & C weighting networks are conformity to standards. * 0.5" standard out size of the microphone. * Time weighting (Fast & Slow) dynamic characteristic modes. * AC / DC output for system expansion.
Page 4
Frequency 31.5 to 8,000 Hz. Microphone type Electric condenser microphone. Microphone size Out size, 12.7 mm DIA. ( 0.5 inch). Weighting Network Characteristics of A & C. Range selector 3 ranges ( 30 to 80 dB, 50 to 100 dB, 80 to 130 dB ).
4. MEASURING PROCEDURE 1. Slide the " A/C weighting selector " ( 3-4 ) to " A " or " C " position for sound level measuring. Note : a. The characteristic table of A, C weighting, please ref. page 7. b.
5. MEASURING CONSIDERATION 1. Please should select the proper weighting selector ( A weighting or C weighting ). Typically the A weighting selector will be engaged. 2. Please select proper measurements range to minimize the tolerance of readout. 3. Please don't keep or operate the instrument at high temperature &...
8. CALIBRATION The sound level meter is built in the internal " Calibration VR " ( 3-10 ) on the front panel. Please according the following procedures to calibrate the instrument accurately, if it is necessary. 1. Prepare the optional " SOUND CALIBRATOR, model: SC-941 or SC-942 ".
9. FREQUENCY WEIGHTING CHARACTERISTICS OF A & C NETWORKS Frequency A Weighting C Weighting Tolerance Charac. Charac. (IEC 61672 class 2) 31.5 -39.4 dB -3 dB ± 3.5 dB -26.2 dB -0.8 dB ± 2.5 dB -16.1 dB -0.2 dB ±...