Welch Allyn Spot VS100 Directions For Use Manual page 48

Vision screener
Hide thumbs Also See for Spot VS100:
Table of Contents

Advertisement

44
Guidance and manufacturer's declaration
Test specifications for enclosure port immunity to RF wireless communications equipment
a
Test frequency
Band
(MHz)
MHz
385
380 - 390
450
430 - 470
710
704 - 787
745
780
810
800 - 960
870
930
1720
1700 - 1990
1845
1970
2450
2400 - 2570
5240
5100 - 5800
5500
5785
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the monitor may be reduced to 1
m. The 1 m test distance is permitted by IEC 61000-4-3.
a
For some services, only the uplink frequencies are included.
b
The carrier shall be modulated using a 50 percent duty cycle square wave signal.
c
As an alternative to FM modulation, 50 percent pulse modulation at 18 Hz may be used because while it does not represent actual
modulation, it would be worst case.
a
Service
Modulation
TETRA 400
Pulse modulation
18 Hz
c
GMRS 460,
FM
±5 kHz
FRS 460
deviation
1 kHz sine
LTE band 13, 17
Pulse modulation
217 Hz
GSM 800/900,
Pulse modulation
TETRA 800,
18 Hz
iDEN 820,
CDMA 850,
LTE Band 5
GSM 1800;
Pulse modulation
CDMA 1900;
217 Hz
GSM 1900;
DECT; LTE Band
1, 3, 4, 25;
UMTS
Bluetooth,
Pulse modulation
WLAN, 802.11
217 Hz
b/g/n, RFID
2450, LTE Band 7
WLAN 802.11
Pulse modulation
a/n
217 Hz
b
Maximum power
Distance (m)
(W)
b
1.8
0.3
2
0.3
b
0.2
0.3
b
2
0.3
b
2
0.3
b
2
0.3
b
0.2
0.3
Welch Allyn Spot Vision Screener
Immunity
test level (V/m)
27
28
9
28
28
28
9

Hide quick links:

Advertisement

Table of Contents
loading