28
Vision Screener operation

Vision Screener criteria

Spot Vision Screener's criteria was developed in accord with the American Academy of
Pediatric Ophthalmology and Strabimus (AAPOS) and American Academy of Pediatrics
(AAP) recommendations for instrument based vision screening. The criteria is age specific
to assist with early detection of amblyopic risk factors in children.
Note
To modify current screening criteria:
1. Perform an export and insert the USB drive into your PC.
2. Go to the USB drive folder and open the Spot folder that you just exported; select the
import folder.
3. Copy the file "SpotCriteria.csv" to the top folder on the USB drive (the file cannot be
within a folder). Once copied, open the file.
4. Make any desired changes using Microsoft Excel or compatible program.
5. Save file as a Comma Separated Value (.csv) file. You must make sure the file is called
"SpotCriteria.csv". Eject the USB drive from your PC and insert into the vision
screener.
6. Go to the Tools menu and select Import/Export. Select the Import button to import
the updated criteria file.
7. To verify the file was imported successfully, go to the Tools menu and select Criteria;
review new settings.
Note

Subjects

The vision screener gives you the ability to populate a subject list for queued subjects.
1. Perform an export, once complete, remove the USB drive from the vision screener
and insert it into your PC.
2. Select the Spot folder that was just created.
3. Select the "db" folder, then select the "SpotSubjects.csv" file.
4. The SpotSubjects.csv file contains six columns; do not change any of the column
headings. Simply add information to the columns as shown below:
Note
Any deviation from this criteria may lead to inaccurate readings.
When custom criteria has been imported and is in use, there will also be a
Restore button at the bottom of the screen. This button will prompt the user to
confirm they want to revert to the default criteria that was shipped with the
device. By confirming, the custom criteria will be deleted from the vision
screener and replaced with the default criteria.
You may add a column before the Patient ID column and call it Location if you
need to create specific locations for each subject.
Patient ID — You can use any combination of alpha, numeric characters and
symbols. This field is used for quick subject look-ups.
First Name — Enter the first name. Use any combination of alpha characters.
Welch Allyn Spot Vision Screener