Appendix For Scale Technicians
Calibration, Single Point
1) Access The Calibration Menu
- Press and hold ZERO key to access settings mode.
P will display, then C will display, followed by AC ?.
- Enter access code: 0000, press ENT. SEL.NNN will display.
2) Calibrate the Indicator
NOTE: If capacity settings are required, refer to Initial
Setup section on page A-1.
a) Remove all weight from the weighing platform.
b) Navigate to SEL.CL1 (if required) using the TARE RECALL
key.
c) Press the GROSS/NET key to proceed. C will flash and
0 lb (or kg) will display.
NOTE: If a lower case c flashes, multi-point calibration
is enabled. If more than one calibration point is desired,
proceed directly to Step 2 on the following page.
d) Press the ZERO key to perform a zero calibration.
e) Load the weighing platform (30% of capacity minimum).
f) On the keypad, enter the weight that is on the platform,
then press the ENT key. The display will show dashes,
then the calibrated weight will display.
g) Press ENT to enter the calibration settings. SEL.CL1 will
display.
h) Press ENT to save settings and exit the settings mode.
SAVE ? and nO will flash. To save calibration settings,
press TARE RECALL to change nO to YES, then press
ENT.
i) At the prompt, enter the 0000 access code, then press
ENT.
j) Calibration is complete.