Apera Instruments PC60 Instruction Manual

Apera Instruments PC60 Instruction Manual

Premium multi-parameter tester ph/ec/tds/salinity/temp.
Hide thumbs Also See for PC60:

Advertisement

PC60 Premium Multi-Parameter Tester
(pH/EC/TDS/Salinity/Temp.)
Instruction Manual
APERA INSTRUMENTS, LLC
www.aperainst.com

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the PC60 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for Apera Instruments PC60

  • Page 1 PC60 Premium Multi-Parameter Tester (pH/EC/TDS/Salinity/Temp.) Instruction Manual APERA INSTRUMENTS, LLC www.aperainst.com...
  • Page 2: Table Of Contents

    Apera Instruments reserves the rights to make any updates of this instruction manual without giving prior notices.
  • Page 3: Battery Installation

    1. Battery Installation Please install batteries according to the following steps. Please note direction of batteries: All POSITIVE SIDES (“+”) FACING UP. (Wrong installation of batteries will cause damage to the tester and potential hazards) ① Pull the battery cap up ②...
  • Page 4: Complete Kit

    3.Complete Kit Diagram - 2 4.Preparation Before Use If it is first-time use or the tester hasn’t been used for over 30 days, pour a little 3M KCl storage solution in the probe cap (about a fifth of the probe cap), and soak the probe for about 5 to 15 minutes.
  • Page 5: Ph Calibration

    5.pH Calibration 5.1 Rinse the probe in pure water and dry it. 5.2 Pour certain amount (about half volume of the calibration bottle) of pH 7.00 and pH 4.00 buffer solution in separate calibration bottles; 5.3 Long press to enter calibration mode; Short press to exit.
  • Page 6: Ph Measurement

    b) For pH Calibration buffer solutions, we recommend that users replace new buffer solution after 10 to 15 times of use to keep the standard buffer’s accuracy. Do NOT pour the used calibration solutions back into the solution bottles in case of contamination. c) pH probes will NOT give accurate and stable readings in distilled or deionized water.
  • Page 7: Conductivity Measurement

    calibration. If it is inaccurate, follow the steps in 7.1 to 7.4 to complete the 2 point of calibration using 12.88 mS buffer solution. 8.Conductivity Measurement Press key to turn on the tester. Rinse the probe in distilled water and dry it. Dip the probe in sample solution, stir gently, and allow it to stand still in the solution until a stable reading is reached.
  • Page 8: Parameter Setting

    e) TDS and conductivity is linear related, and its conversion factor is 0.40-1.00. Adjust the factor in parameter setting P5. The factory default setting is 0.71. Salinity and conductivity is linear related, and its conversion factor is 0.5. The tester only needs to be calibrated in Conductivity mode, then after calibration of conductivity, the meter can switch from conductivity to TDS or salinity.
  • Page 9: Technical Specifications

    9.3 Parameter Setting Instruction a) Select standard pH buffer solution (P1): There are two options of standard buffer solutions: USA series and NIST series. Refer to following chart: pH Standard Buffer Solution Series Icons USA series NIST series 1.68 pH and 4.00 pH 1.68 pH and 4.01 pH Three-point 7.00 pH...
  • Page 10: Tester Functions

    Screw off the probe ring, unplug the probe, plug in the new replacement probe (pay attention to the probe’s position), and screw on the probe ring. The model numbers of replacement probes that are compatible with PC60 are:  PC60-E ...
  • Page 11: Warranty

    We warrant this instrument to be free from defects in material and workmanship and agree to repair or replace free of charge, at option of APERA INSTRUMENTS, LLC, any malfunctioned or damaged product attributable to responsibility of APERA INSTRUMENTS, LLC for a period of two years from the delivery (a six-month limited warranty applies to probes).
  • Page 12 APERA INSTRUMENTS, LLC Address: 977 Old Henderson Rd, Columbus Ohio 43220 Tel: 1-614-285-3080 Email: info@aperainst.com Website: www.aperainst.com...